Enhanced Hemocompatibility via Bivalirudin and Bicarbonate as Alternative to Heparin in a Catheter‐Based Axial‐Flow System
Catheterization and Cardiovascular Interventions · 2025
DOI: 10.1002/ccd.31663 ↗At a glance
Vergleichsstudie in Catheterization and Cardiovascular Interventions, 2025. Written by Kaitlyn R Ammann and 7 co-authors. Institution: Arizona Center for Accelerated Biomedical Innovation, Sarver Heart Center, University of Arizona Health Sciences, The University of Arizona, Tucson, …. Major MEDLINE topics: Hirudins, Heparin, Peptide Fragments, Heart-Assist Devices. Cited 1 times so far (Crossref); 48 references. The full text is freely available (open access, CC BY-NC-ND 4.0). Funded by: This work was supported by an unrestricted educational grant from Johnson and Johnson/Abiomed Inc as well as from a grant from the Arizona Center for Accelerat….
Abstract (original language)
Background
Cardiovascular therapeutic devices typically require systemic heparin due to underlying thrombotic risk. The Impella axial flow system further relies on internal perfusion with either a heparin‐containing or sodium bicarbonate purge solution during operation. The combination of systemic and device‐mediated heparin often contributes to varying levels of overall anticoagulation, raising the possibility of increased bleeding risk and complications due to heparin sensitivity. To reduce adverse events, efforts have been ongoing to eliminate heparin, both systemically and via the device. Here, we investigate bivalirudin as an alternative to systemic heparin and sodium bicarbonate as an alternative to local device heparin in the purge solution.
Methods
We examined hemocompatibility in a mock loop environment with circulating whole porcine blood to determine feasibility of heparin‐free Impella system. Anticoagulated (heparin or bivalirudin) blood was loaded into a circulatory loop with a contained Impella 5.5 and run for 4 h with either heparin or sodium bicarbonate‐containing purge solution. Blood samples were collected serially and analyzed for metrics of hemocompatibility: hemolysis, platelet activation, and vWF degradation.
Results
Bivalirudin and sodium bicarbonate adjunctive environment led to similar or improved hemocompatibility as compared to heparin‐containing environment—notably, decreased hemolysis (40 ± 20.8 mg/dL average decrease) and preserved platelet activity measured by P‐selectin exposure (8.0 ± 5.6% average increase).
Conclusions
Our data supports the potential to create an overall heparin‐free approach simplifying anticoagulation management, offering potential for overall reduced bleeding risk and heparin‐related complications.
Subjects
MeSH terms (MEDLINE)
Author keywords
Substances
Authors and institutions
- Arizona Center for Accelerated Biomedical Innovation, Sarver Heart Center, University of Arizona Health Sciences, The University of Arizona, Tucson, Arizona, USA.
- Abiomed Inc., Danvers, Massachusetts, USA.
- Department of Medicine, Department of Biomedical Engineering, Arizona Center for Accelerated Biomedical Innovation, Sarver Heart Center, The University of Arizona, Tucson, Arizona, USA.
ORCID profiles: Kaitlyn R Ammann
Funding
- This work was supported by an unrestricted educational grant from Johnson and Johnson/Abiomed Inc as well as from a grant from the Arizona Center for Accelerat…
Cite
APA
Ammann, K. R., Outridge, C. E., Silver, T., Muslmani, S., Ding, J., Gilman, V., Corbett, S., & Slepian, M. J. (2025). Enhanced Hemocompatibility via Bivalirudin and Bicarbonate as Alternative to Heparin in a Catheter‐Based Axial‐Flow System. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31663
Vancouver
Ammann KR, Outridge CE, Silver T, Muslmani S, Ding J, Gilman V, et al. Enhanced Hemocompatibility via Bivalirudin and Bicarbonate as Alternative to Heparin in a Catheter‐Based Axial‐Flow System. Catheterization and Cardiovascular Interventions. 2025. doi:10.1002/ccd.31663