EUVIMEDEUVIMED entdecken
Systematische ÜbersichtMetaanalyse

Combination of Suture‐Plug Vascular Closure Devices Versus Suture‐Only Devices Following Transcatheter Aortic Valve Implantation: A Systematic Review and Meta‐Analysis

Saddik Haddad iD, George Bashour iD, Mamas A Mamas iD

Catheterization and Cardiovascular Interventions · 2025

DOI: 10.1002/ccd.31655 ↗

Kurz gefasst

Systematische Übersicht in der Zeitschrift Catheterization and Cardiovascular Interventions, 2025. Verfasst von Saddik Haddad und 2 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Medicine, Tishreen University, Latakia, Syria. Hauptthemen laut MEDLINE: Transcatheter Aortic Valve Replacement, Vascular Closure Devices, Hemostatic Techniques, Aortic Valve Stenosis. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 22 Literaturangaben. Gefördert von: The authors received no specific funding for this work..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Transcatheter aortic valve implantation (TAVI) is an effective and safe intervention for severe symptomatic aortic stenosis regardless of surgical risk. Vascular complications following TAVI are considered among the most common and serious complications, with a higher risk of mortality.

Aims

We aimed to perform a systematic review and meta‐analysis to evaluate the safety and efficacy of the combination of suture‐collagen plug vascular closure devices (VCDs) compared with suture‐only devices.

Methods

We have systematically searched the literature from inception until January 4, 2025, for cohort studies and clinical trials comparing the combination of suture‐collagen plug vascular closure devices (VCDs) versus suture‐only VCDs.

Results

A total of 2064 patients across six studies were included in the analysis. All studies compared dual Perclose ProGlide versus 1 Perclose ProGlide + 1 Angio‐Seal, except one study that used either one Prostar or dual ProGlide in the suture group. The combination group of suture‐plug VCDs was associated with a statistically significant reduction of VCDs failure (RR = 0.14, CI 95% [0.09–0.23], p < 0.0001, I 2 = 58.6938%), bleeding complications (RR = 0.35, CI95% [0.24–0.53], p < 0.0001, 0.0%), major vascular complications (RR = 0.42, CI 95% [0.28, 0.63], p < 0.0001, I 2 = 50.4%), and minor vascular complications (RR = 0.54, CI 95% [0.43, 0.66], p < 0.0001, I 2 = 0%). On the other hand, all‐cause mortality did not have a statistically significant difference between the two groups (RR = 0.54, CI 95% [0.28, 1.05], p = 0.0709, I 2 = 0%).

Conclusions

In patients undergoing TAVI, the combination strategy of suture‐plug VCDs is superior to suture‐only devices regarding major and minor vascular complications, bleeding complications, and VCD failure. The significantly higher VCD failure of the suture‐only group observed in our study led to the use of additional closure devices to secure hemostasis, raising concerns about a higher risk of peripheral ischemia.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Transcatheter Aortic Valve ReplacementVascular Closure DevicesHemostatic TechniquesAortic Valve StenosisHemorrhageSuturesPostoperative HemorrhageHumansTreatment OutcomeRisk FactorsAged, 80 and overAgedMaleFemaleRisk AssessmentEquipment DesignPunctures

Stichwörter der Autor:innen

TAVIsuture vascular closure devicessuture‐collagen plug vascular closure devicestranscatheter aortic valve implantation

Autor:innen und Institutionen

  • Faculty of Medicine, Tishreen University, Latakia, Syria.
  • Cancer Research Center, Tishreen University, Latakia, Syria.
  • Keele Cardiovascular Research Group, Keele University, Keele, UK.

ORCID-Profile: Saddik Haddad, George Bashour, Mamas A Mamas

Förderung

  • The authors received no specific funding for this work.

Zitieren

APA

Haddad, S., Bashour, G., & Mamas, M. A. (2025). Combination of Suture‐Plug Vascular Closure Devices Versus Suture‐Only Devices Following Transcatheter Aortic Valve Implantation: A Systematic Review and Meta‐Analysis. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31655

Vancouver

Haddad S, Bashour G, Mamas MA. Combination of Suture‐Plug Vascular Closure Devices Versus Suture‐Only Devices Following Transcatheter Aortic Valve Implantation: A Systematic Review and Meta‐Analysis. Catheterization and Cardiovascular Interventions. 2025. doi:10.1002/ccd.31655

BibTeX
@article{Haddad2025,
  author = {Haddad, Saddik and Bashour, George and Mamas, Mamas A},
  title = {Combination of Suture‐Plug Vascular Closure Devices Versus Suture‐Only Devices Following Transcatheter Aortic Valve Implantation: A Systematic Review and Meta‐Analysis},
  journal = {Catheterization and Cardiovascular Interventions},