Beobachtungsstudie in der Zeitschrift Catheterization and Cardiovascular Interventions, online erschienen am 19.04.2024. Verfasst von Fortunato Iacovelli und 13 weiteren Autor:innen. Beteiligte Einrichtung: Division of University Cardiology, Cardiothoracic Department, Policlinico University Hospital, Bari, Italy. Hauptthemen laut MEDLINE: Femoral Artery, Transcatheter Aortic Valve Replacement, Catheterization, Peripheral, Hemostatic Techniques. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 20 Literaturangaben.
Abstract (Originalsprache)
Background
Despite the use of two crossed Perclose ProGlide™ (Abbott Vascular Devices) is the most widespread technique to close the main arterial access in transfemoral transcatheter aortic valve implantation (TF‐TAVI), the safest and most effective strategy still remains much debated.
Aims
The aim of the present study was to evaluate the performance of a single Perclose ProGlide suture‐mediated closure device to obtain femoral hemostasis after sheathless implantation of self‐expanding transcatheter heart valves through their 14 F‐equivalent fix delivery systems.
Methods
This prospective observational study included 439 patients undergoing TF‐TAVI at the “Montevergine” Clinic of Mercogliano, Italy. All patients underwent hemostasis of the large‐bore access using a single Perclose ProGlide with preclose technique, after sheathless implantation of self‐expanding transcatheter heart valves through 14 F‐equivalent fix delivery systems. A multidetector computed tomography analysis of size, tortuosity, atherosclerotic, and calcification burdens of the ilio‐femoral access route was made by a dedicated corelab. Vascular complications (VCs), percutaneous closure device (PCD) failure, and bleedings were adjudicated by a clinical events committee.
Results
A total of 81 different VCs were observed in 60 patients (13.7%); among these, 41 (5% of patients) were categorized as major. PCD failure occurred in 14 patients (3.2%). At the logistic regression analysis, no predictors of PCD failure have been identified.
Conclusion
This registry suggests that the use of a single suture‐mediated closure device could be considered a safe and efficient technique to achieve access site hemostasis in patients undergoing TF‐TAVI through 14 F‐equivalent fix delivery systems.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of University Cardiology, Cardiothoracic Department, Policlinico University Hospital, Bari, Italy.
- Division of Cardiology, "SS. Annunziata" Hospital, Taranto, Italy.
- Clinical Trials Unit, School of Medicine, University of Bern, Bern, Switzerland.
- School of Medicine, University of Basilicata, Potenza, Italy.
- Division of Cardiology, "Papardo" Hospital, Messina, Italy.
- Interventional Cardiology Service, "Montevergine" Clinic, GVM Care & Research, Mercogliano, Italy.
ORCID-Profile: Fortunato Iacovelli, Osvaldo Burattini, Francesca Sturdà, Mattia Branca, Eugenio Stabile, Luigi Fimiani, Angelo Cioppa, Alessandro Cafaro
Zitieren
APA
Iacovelli, F., Burattini, O., Sturdà, F., Branca, M., Stabile, E., Fimiani, L., Salemme, L., Cioppa, A., Pucciarelli, A., Cafaro, A., Bortone, A. S., Contegiacomo, G., Pignatelli, A., & Tesorio, T. (2024). Single suture‐mediated closure system after transfemoral transcatheter aortic valve implantation: A single‐center real‐world experience. Catheterization and Cardiovascular Interventions, 103(7), 1125-1137. https://doi.org/10.1002/ccd.31054
Vancouver
Iacovelli F, Burattini O, Sturdà F, Branca M, Stabile E, Fimiani L, et al. Single suture‐mediated closure system after transfemoral transcatheter aortic valve implantation: A single‐center real‐world experience. Catheterization and Cardiovascular Interventions. 2024;103(7):1125-1137. doi:10.1002/ccd.31054
BibTeX
@article{Iacovelli2024,
author = {Iacovelli, Fortunato and Burattini, Osvaldo and Sturdà, Francesca and Branca, Mattia and Stabile, Eugenio and Fimiani, Luigi and Salemme, Luigi and Cioppa, Angelo and Pucciarelli, Armando and Cafaro, Alessandro and Bortone, Alessandro Santo and Contegiacomo, Gaetano and Pignatelli, Antonio and Tesorio, Tullio},
title = {Single suture‐mediated closure system after transfemoral transcatheter aortic valve implantation: A single‐center real‐world experience},
journal = {Catheterization and Cardiovascular Interventions},
year = {2024},
volume = {103},
number = {7},
pages = {1125--1137},
doi = {10.1002/ccd.31054}
}
EUVIMED