Accurate and Safe Directional Coronary Atherectomy Using the X‐Ray Detector's Calculated Optimal Angular View
Catheterization and Cardiovascular Interventions · 2025
DOI: 10.1002/ccd.31566 ↗Kurz gefasst
Fachartikel in der Zeitschrift Catheterization and Cardiovascular Interventions, 2025. Verfasst von Toru Takii und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Sendai Open Hospital, Sendai, Japan. Hauptthemen laut MEDLINE: Atherectomy, Coronary, Coronary Angiography, Coronary Artery Disease, Coronary Vessels. Die Arbeit nennt 9 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: The authors received no specific funding for this work..
Abstract (Originalsprache)
Background
Directional coronary atherectomy (DCA) is effective for bifurcation lesions but carries a risk of coronary perforation, due to geographical and recognition errors during the procedure. Geographical errors involve incorrect targeting of plaque‐free areas, whereas recognition errors stem from the misinterpretation of visual information during catheter alignment. To minimize these risks and improve procedural safety and efficacy, it is crucial to align the DCA catheter accurately with the target plaque by utilizing the morphological characteristics of the housing.
Aims
This study aimed to determine whether visual recognition of DCA housing from the front or side, as opposed to an oblique angle, reduces cutting errors.
Methods
Fifteen physicians and five medical staff members performed a bench test in which a DCA catheter was rotated 0° (frontal view), 45° clockwise (oblique view), and 90° clockwise (lateral view) under fluoroscopy. The error angles at each position were compared ( p < 0.05).
Results
Error angles were 4.4 ± 3.4° (frontal view), 10.1 ± 4.6° (lateral view), and 13.6 ± 7.5° (oblique view). The frontal view showed fewer errors than the lateral and oblique views ( p = 0.0016 and p < 0.0001, respectively). No significant differences were observed between the physicians and medical staff.
Conclusion
Recognizing DCA housing from the frontal or lateral view calculated from intravascular ultrasound (IVUS) significantly reduces recognition errors in plaque cutting, enables safer and more effective procedures, even for less experienced operators, and has the potential to improve the safety and outcomes of DCA procedures.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Sendai Open Hospital, Sendai, Japan.
- Department of Clinical Engineering, Sendai Open Hospital, Sendai, Japan.
ORCID-Profile: Toru Takii
Förderung
- The authors received no specific funding for this work.
Zitieren
APA
Takii, T., Tazawa, M., Tanita, A., Sunamura, S., Ogata, T., Noda, K., & Namiuchi, S. (2025). Accurate and Safe Directional Coronary Atherectomy Using the X‐Ray Detector's Calculated Optimal Angular View. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31566
Vancouver
Takii T, Tazawa M, Tanita A, Sunamura S, Ogata T, Noda K, et al. Accurate and Safe Directional Coronary Atherectomy Using the X‐Ray Detector's Calculated Optimal Angular View. Catheterization and Cardiovascular Interventions. 2025. doi:10.1002/ccd.31566
BibTeX
@article{Takii2025,
author = {Takii, Toru and Tazawa, Mimori and Tanita, Atsushi and Sunamura, Shinichiro and Ogata, Tsuyoshi and Noda, Kazuki and Namiuchi, Shigeto},
title = {Accurate and Safe Directional Coronary Atherectomy Using the X‐Ray Detector's Calculated Optimal Angular View},
journal = {Catheterization and Cardiovascular Interventions},
year = {2025},
doi = {10.1002/ccd.31566}
}