Image‐Derived Modeling to Assess Coronary Proximity in Patients Undergoing Transcatheter Pulmonary Valve Replacement With Self‐Expanding Valves
Catheterization and Cardiovascular Interventions · 2025
DOI: 10.1002/ccd.31469 ↗At a glance
Journal article in Catheterization and Cardiovascular Interventions, 2025. Written by Yuval Barak-Corren and 8 co-authors. Institution: Department of Pediatrics, Division of Pediatric Cardiology, The Children's Hospital of Philadelphia, Philadelphia, Pennsylvania, USA. Major MEDLINE topics: Pulmonary Valve, Cardiac Catheterization, Heart Valve Prosthesis Implantation, Heart Valve Prosthesis. Cited 1 times so far (Crossref); 22 references. The full text is freely available in PubMed Central. Funded by: NHLBI NIH HHS.
Abstract (original language)
Background
The potential for coronary artery compression (CC) during transcatheter pulmonary valve replacement (TPVR) using self‐expanding valves (SEV) is not fully understood, yet anecdotal reports suggest that this risk exists. Aims and Methods We performed a retrospective cohort study of patients evaluated for SEV‐TPVR to evaluate the relationship between the right ventricular outflow tract (RVOT) and coronary arteries (CA). CT‐derived segmentations of the RVOT and CA were created using machine learning. A 2D map of the distance between the RVOT surface and CA, in systole and diastole, was created. In the subset of patients with post‐procedural CTA, the distance before and after TPVR was measured.
Results
Forty‐two individuals underwent screening for SEV‐TPVR, of which 83% ( n = 35) had SEV implanted (Harmony = 24; Alterra = 11). Median age was 22.9 years (range 12−60) and 76% had tetralogy of Fallot (TOF). There was no significant change in the distance between the RVOT and LCA between diastole and systole ( p = 0.31), yet the RVOT area nearest to the LCA displaced proximally by 11 mm (IQR: 5.6−19.9) in systole. In 8 patients with pre‐ and post‐TPVR CTA, no statistically significant differences were observed in the RVOT‐to‐LCA relation after intervention. The distance to the LCA was smaller in pulmonary stenosis/atresia patients than those with TOF (median distance 1.2 and 2.1 mm, respectively; p = 0.185).
Conclusion
The RVOT area in closest proximity to LCA is dynamic and should be considered when planning TPVR. Special attention should be given to patients with a diagnosis of pulmonary stenosis/atresia.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Pediatrics, Division of Pediatric Cardiology, The Children's Hospital of Philadelphia, Philadelphia, Pennsylvania, USA.
- Division of Pediatric Cardiology, The Mount Sinai Hospital, New York, New York, USA.
- Hospital of the University of Pennsylvania, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
- Department of Anesthesiology and Critical Care Medicine, The Children's Hospital of Philadelphia, Philadelphia, Pennsylvania, USA.
- Laboratory for Percutaneous Surgery, Queen's University, Kingston, Ontario, CA.
ORCID profiles: Mudit Gupta, Michael L O'Byrne, Matthew A Jolley
Funding
- NHLBI NIH HHS
Cite
APA
Barak-Corren, Y., Obsekov, V., Gupta, M., Herz, C., Amin, S., Lasso, A., O'Byrne, M. L., Gillespie, M. J., & Jolley, M. A. (2025). Image‐Derived Modeling to Assess Coronary Proximity in Patients Undergoing Transcatheter Pulmonary Valve Replacement With Self‐Expanding Valves. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31469
Vancouver
Barak-Corren Y, Obsekov V, Gupta M, Herz C, Amin S, Lasso A, et al. Image‐Derived Modeling to Assess Coronary Proximity in Patients Undergoing Transcatheter Pulmonary Valve Replacement With Self‐Expanding Valves. Catheterization and Cardiovascular Interventions. 2025. doi:10.1002/ccd.31469
BibTeX
@article{BarakCorren2025,
author = {Barak-Corren, Yuval and Obsekov, Vladislav and Gupta, Mudit and Herz, Christian and Amin, Silvani and Lasso, Andras and O'Byrne, Michael L and Gillespie, Matthew J and Jolley, Matthew A},
title = {Image‐Derived Modeling to Assess Coronary Proximity in Patients Undergoing Transcatheter Pulmonary Valve Replacement With Self‐Expanding Valves},
journal = {Catheterization and Cardiovascular Interventions},
year = {2025},
doi = {10.1002/ccd.31469}
}
EUVIMED