EUVIMEDSkopri EUVIMED
ValidierungsstudieVergleichsstudie

Machine Learning in Intravascular Ultrasound: Validating Automated Lesion Assessment for Complex Coronary Interventions

Jason Galo iD, Abhishek Chaturvedi, Abdullah Al-Qaraghuli, Pablo M Rubio, Cheng Zhang, Yirga Kahsay, Beni Rai Verma, Kalyan R Chitturi, Gary S Mintz, Ron Waksman iD, Brian C Case, Hayder Hashim and 1 more

Catheterization and Cardiovascular Interventions · 2025-05 · 105 (6) · 1320-1328

DOI: 10.1002/ccd.31458 ↗

At a glance

Validierungsstudie in Catheterization and Cardiovascular Interventions, published online on 2025-02-21. Written by Jason Galo and 12 co-authors. Institution: Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, District of Columbia, USA. Major MEDLINE topics: Ultrasonography, Interventional, Machine Learning, Coronary Artery Disease, Coronary Vessels. Cited 3 times so far (Crossref); 13 references.

AstrattSubjectsDettalji biblijografiċiCiteRelated studies

Abstract (original language)

Background

Intravascular ultrasound (IVUS) is essential for assessing complex coronary lesions, but remains underutilized in part due to difficulties in image interpretation. The AVVIGO IVUS Automated Lesion Assessment (ALA) software, which uses machine learning (ML) for automatic segmentation, promises to simplify lesion assessment. This study evaluated the agreement in stent size selection between ALA, an independent core laboratory (CL), and an expert interventional cardiologist (IC) for complex lesions.

Aims

The primary endpoint was the agreement in stent size selection, within 0.25 mm, of AVVIGO ALA automatic segmentation of Class I lesions against the gold‐standard measurement by an independent CL analysis and against an expert IC, (H. H.). The secondary endpoint was to assess the relative differences between AVVIGO ALA and CL, AVVIGO ALA and IC, and CL and IC, in vessel and lumen areas.

Methods

Patients with complex coronary lesions, including left main bifurcation, long, and severely calcified lesions, were retrospectively analyzed using IVUS with ALA. Stent size selection and area measurements by ALA were compared against a CL and IC using established sizing methods.

Results

In 48 patients, ALA demonstrated high agreement with CL (92%–100%) and IC (91%–98.5%) in stent size selection across lesion subtypes using recommended sizing methods. Lumen‐based sizing achieved higher agreement than vessel‐based sizing, particularly in calcified lesions (100% vs. 87%). The variability in relative difference in measurements between ALA and CL was greater than IC and CL in distal vessel and lesion vessel areas. The relative difference in measurements between ALA and IC was greater in vessel‐based sizing compared to lumen‐based sizing in the distal reference marker.

Conclusion

AVVIGO ALA demonstrated high agreement in stent size selection compared to a CL and expert IC. ML's ability to automate IVUS analysis may improve operator efficiency, reduce radiation exposure, and enhance the adoption of intravascular imaging in routine practice. It remains to be seen if it will impact of adoption of IVUS to guide complex PCI.

Subjects

MeSH terms (MEDLINE)

Ultrasonography, InterventionalMachine LearningCoronary Artery DiseaseCoronary VesselsPercutaneous Coronary InterventionImage Interpretation, Computer-AssistedHumansPredictive Value of TestsFemaleMaleRetrospective StudiesReproducibility of ResultsMiddle AgedStentsAgedAutomationProsthesis DesignSeverity of Illness IndexObserver VariationTreatment Outcome

Author keywords

Automated Lesion Assessment (ALA)calcified lesionscomplex PCIcoronary lesionsintravascular ultrasound (IVUS)lumen‐based sizingmachine learning (ML)stent size selectionvessel‐based sizing

Authors and institutions

  • Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, District of Columbia, USA.
  • Department of Internal Medicine, MedStar Washington Hospital Center, Washington, District of Columbia, USA.
  • Cardiovascular Research Foundation, New York, New York, USA.

ORCID profiles: Jason Galo, Ron Waksman

Cite

APA

Galo, J., Chaturvedi, A., Al-Qaraghuli, A., Rubio, P. M., Zhang, C., Kahsay, Y., Verma, B. R., Chitturi, K. R., Mintz, G. S., Waksman, R., Case, B. C., Hashim, H., & Garcia-Garcia, H. M. (2025). Machine Learning in Intravascular Ultrasound: Validating Automated Lesion Assessment for Complex Coronary Interventions. Catheterization and Cardiovascular Interventions, 105(6), 1320-1328. https://doi.org/10.1002/ccd.31458

Vancouver

Galo J, Chaturvedi A, Al-Qaraghuli A, Rubio PM, Zhang C, Kahsay Y, et al. Machine Learning in Intravascular Ultrasound: Validating Automated Lesion Assessment for Complex Coronary Interventions. Catheterization and Cardiovascular Interventions. 2025;105(6):1320-1328. doi:10.1002/ccd.31458

BibTeX
@article{Galo2025,
  author = {Galo, Jason and Chaturvedi, Abhishek and Al-Qaraghuli, Abdullah and Rubio, Pablo M and Zhang, Cheng and Kahsay, Yirga and Verma, Beni Rai and Chitturi, Kalyan R and Mintz, Gary S and Waksman, Ron and Case, Brian C and Hashim, Hayder and Garcia-Garcia, Hector M},
  title = {Machine Learning in Intravascular Ultrasound: Validating Automated Lesion Assessment for Complex Coronary Interventions},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2025},
  volume = {105},
  number = {6},
  pages = {1320--1328},
  doi = {10.1002/ccd.31458}
}
Start for free →