EUVIMEDOpdag EUVIMED
VergleichsstudieMultizentrische Studie

Physiological Performance of Drug‐Coated Balloons in Small Coronary Arteries PICCOLETO II μFR

Simone Fezzi iD, Aurora Trevisanello, Dario Buccheri, Marco Borgi, Pedro Silva Orrego, Giuseppe Biondi Zoccai, Bernardo Cortese iD

Catheterization and Cardiovascular Interventions · 2025-02 · 105 (3) · 643-649

DOI: 10.1002/ccd.31376 ↗

At a glance

Vergleichsstudie in Catheterization and Cardiovascular Interventions, published online on 2024-12-24. Written by Simone Fezzi and 6 co-authors. Institution: Department of Medicine, Division of Cardiology, University of Verona, Verona, Italy. Major MEDLINE topics: Angioplasty, Balloon, Coronary, Cardiac Catheters, Cardiovascular Agents, Coated Materials, Biocompatible. Cited 6 times so far (Crossref); 17 references.

AbstractSubjectsBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Drug‐coated balloons (DCB) are emerging as an alternative to permanent implants for managing de novo coronary artery disease, particularly in small vessels (SVD). This sub‐analysis of the PICCOLETO II study aimed to compare the performance of DCB and DES in terms of Murray's law‐based quantitative flow ratio (μFR) changes between baseline, post‐percutaneous coronary intervention (PCI), and follow‐up.

Methods

Patients with a clinical indication for PCI were assigned to receive either Xience DES or Elutax SV/Emperor DCB. Coronary angiograms were blindly analyzed by an independent Core Laboratory (Consorzio Futuro in Ricerca, University of Ferrara, Italy).

Results

Among 232 patients, 59 were included in this analysis. Pre‐PCI μFR was comparable between groups (0.65 ± 0.29 vs. 0.58 ± 0.25; p = 0.20). Post‐PCI, the DCB group had a lower acute functional gain compared to DES (+0.21 vs. +0.31; p = 0.064), with lower μFR values (0.86 ± 0.36 vs. 0.89 ± 0.3; p = 0.074). At 6‐month follow‐up, μFR values were similar between groups (0.84 ± 0.29 vs. 0.84 ± 0.31, p = 0.93), with a comparable late functional loss (−0.02 vs. −0.05; p = 0.93). Angiographic late lumen loss was significantly lower in the DCB group (−0.03 vs. +0.29 mm; p = 0.027).

Conclusions

DCB‐based PCI showed a trend toward less favorable immediate post‐PCI functional result but a comparable performance at follow‐up, suggesting its potential in treating de novo SVD.

Subjects

MeSH terms (MEDLINE)

Angioplasty, Balloon, CoronaryCardiac CathetersCardiovascular AgentsCoated Materials, BiocompatibleCoronary Artery DiseaseCoronary VesselsDrug-Eluting StentsAgedFemaleHumansMaleMiddle AgedCoronary AngiographyCoronary CirculationEquipment DesignItalyProspective StudiesTime FactorsTreatment Outcome

Substances

Cardiovascular AgentsCoated Materials, Biocompatible

Authors and institutions

  • Department of Medicine, Division of Cardiology, University of Verona, Verona, Italy.
  • Cardiovascular Department, Ospedale di Trapani, Trapani, Italy.
  • Department of Clinical and Experimental Medicine, Policlinico G. Martino, University of Messina, Messina, Italy.
  • Interventional Cardiology, ASST Fatebenefratelli-Sacco, Milano, Italy.
  • Department of Medical-Surgical Sciences and Biotechnologies, Sapienza University of Rome, Latina, Italy.
  • DCB Academy, Milano, Italy.

ORCID profiles: Simone Fezzi, Bernardo Cortese

Cite

APA

Fezzi, S., Trevisanello, A., Buccheri, D., Borgi, M., Orrego, P. S., Zoccai, G. B., & Cortese, B. (2025). Physiological Performance of Drug‐Coated Balloons in Small Coronary Arteries PICCOLETO II μFR. Catheterization and Cardiovascular Interventions, 105(3), 643-649. https://doi.org/10.1002/ccd.31376

Vancouver

Fezzi S, Trevisanello A, Buccheri D, Borgi M, Orrego PS, Zoccai GB, et al. Physiological Performance of Drug‐Coated Balloons in Small Coronary Arteries PICCOLETO II μFR. Catheterization and Cardiovascular Interventions. 2025;105(3):643-649. doi:10.1002/ccd.31376

BibTeX
@article{Fezzi2025,
  author = {Fezzi, Simone and Trevisanello, Aurora and Buccheri, Dario and Borgi, Marco and Orrego, Pedro Silva and Zoccai, Giuseppe Biondi and Cortese, Bernardo},
  title = {Physiological Performance of Drug‐Coated Balloons in Small Coronary Arteries PICCOLETO II μFR},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2025},
  volume = {105},
  number = {3},
  pages = {643--649},
  doi = {10.1002/ccd.31376}
}