Microcatheter tip fracture and acute vessel occlusion rescued with subintimal tracking and re‐entry: Move the tip
Catheterization and Cardiovascular Interventions · 2024
DOI: 10.1002/ccd.31264 ↗At a glance
Fallbericht in Catheterization and Cardiovascular Interventions, 2024. Written by Francesco Giangiacomi and 3 co-authors. Institution: Department of Clinical Sciences and Community Health, University of Milan, Milan, Italy. Major MEDLINE topics: Cardiac Catheters, Coronary Occlusion, Angioplasty, Balloon, Coronary, Equipment Failure. Cited 1 times so far (Crossref); 7 references. The full text is freely available (open access, CC BY 4.0).
Abstract (original language)
The development of microcatheters capable of advancing through long, calcified, and tortuous vessels, produced an increase in the success rate of complex percutaneous coronary interventions. However, the aggressive manipulation of these materials may lead to potentially severe and threatening complications. We present a case of Turnpike Spiral (Teleflex) tip fracture, which caused an acute occlusion of the proximal right coronary artery. The coronary artery was tortuous and had two heavily calcified lesions in the proximal segment, uncrossable to both small angioplasty balloons and standard microcatheters. After the incarceration of the tip of the Turnpike Spiral in the calcium, it got fractured during the attempt of removing it from the vessel. This led to ST‐segment elevation and patient instability. Due to the impossibility of advancing a second guidewire into the true lumen, we decided to proceed to antegrade dissection and re‐entry using the scratch‐and‐go and subintimal tracking and re‐entry technique. We placed a guide‐extension and we used a Gaia II (Asahi Intecc) to enter the subintimal space and advance over the fractured tip. The wire was exchanged for a Gladius MG (Asahi Intecc), which re‐entered the true lumen at the level of the acute margin of the heart. Intravascular ultrasound was used to confirm the re‐entry point, and three drug eluting stents were placed. In this way we rapidly restored the flow in the right coronary artery, and we prevented a myocardial infarction.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Clinical Sciences and Community Health, University of Milan, Milan, Italy.
- Department of Cardio-Thoracic-Vascular Diseases, Foundation IRCCS Ca' Granda Ospedale Maggiore Policlinico, Milan, Italy.
ORCID profiles: Francesco Giangiacomi, Gabriele Tumminello
Cite
APA
Giangiacomi, F., Mircoli, L., Tumminello, G., & Carugo, S. (2024). Microcatheter tip fracture and acute vessel occlusion rescued with subintimal tracking and re‐entry: Move the tip. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31264
Vancouver
Giangiacomi F, Mircoli L, Tumminello G, Carugo S. Microcatheter tip fracture and acute vessel occlusion rescued with subintimal tracking and re‐entry: Move the tip. Catheterization and Cardiovascular Interventions. 2024. doi:10.1002/ccd.31264
BibTeX
@article{Giangiacomi2024,
author = {Giangiacomi, Francesco and Mircoli, Luca and Tumminello, Gabriele and Carugo, Stefano},
title = {Microcatheter tip fracture and acute vessel occlusion rescued with subintimal tracking and re‐entry: Move the tip},
journal = {Catheterization and Cardiovascular Interventions},
year = {2024},
doi = {10.1002/ccd.31264}
}
EUVIMED