Evaluation of microcirculatory protection in percutaneous revascularisation: A stent implantation technique and device comparison
Catheterization and Cardiovascular Interventions · 2024
DOI: 10.1002/ccd.31155 ↗At a glance
Randomisierte kontrollierte Studie in Catheterization and Cardiovascular Interventions, 2024. Written by Muhammad Aetesam-Ur-Rahman and 10 co-authors. Institution: Department of Interventional Cardiology, Royal Papworth Hospital, Cambridge, UK. Major MEDLINE topics: Tomography, Optical Coherence, Microcirculation, Drug-Eluting Stents, Prosthesis Design. 35 references. The full text is freely available (open access, CC BY 4.0). Funded by: Abbott Vascular, Cambridge Biomedical Research Centre, Heart and Lung Research Institute.
Abstract (original language)
Background
Coronary microvascular dysfunction (CMD) after percutaneous coronary intervention (PCI) is prognostically important and may also be a cause of persistent angina. The stent balloon inflation technique or material properties may influence the degree of CMD post‐PCI.
Methods
Thirty‐six patients with stable angina attending for elective PCI were randomized to either slow drug eluting stent (DES) implantation technique (DES slow group): +2 atm. every 5 s., maintained for a further 30 s or a standard stent implantation technique (DES std group): rapid inflation and deflation. PressureWire X with thermodilution at rest and hyperemia and optical coherence tomography (OCT) were performed pre‐ and post‐PCI. Combined primary endpoints were changes in index of microvascular resistance (delta IMR) and coronary flow reserve (delta CFR) following PCI. The secondary endpoints included differences in cardiac troponin I (delta cTnI) at 6 h post‐PCI, Seattle angina questionnaire (SAQ) at 1, 3, 6, and 12 months and OCT measures of stent results immediately post‐PCI and at 3 months.
Results
Both groups were well matched, with similar baseline characteristics and OCT‐defined plaque characteristics. Delta IMR was significantly better in the DES slow PCI arm with a median difference of −4.14 (95% CI ‐10.49, −0.39, p = 0.04 ). Delta CFR was also numerically higher with a median difference of 0.47 (95% CI −0.52, 1.31, p = 0.46). This did not translate to improved delta median cTnI (1.5 (34.8) vs. 0 (27.5) ng/L, p = 0.75) or median SAQ score at 3 months, (85 (20) vs. 95 (17.5), p = 0.47).
Conclusion
Slow stent implantation is associated with less CMD after elective PCI in patients with stable angina.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Department of Interventional Cardiology, Royal Papworth Hospital, Cambridge, UK.
- Division of Cardiovascular Medicine, University of Cambridge, Cambridge, UK.
ORCID profiles: Muhammad Aetesam-Ur-Rahman, Joana Oliveira, Bharat Khialani, Stephen P Hoole
Funding
- Abbott Vascular
- Cambridge Biomedical Research Centre
- Heart and Lung Research Institute
Cite
APA
Aetesam-Ur-Rahman, M., Zhao, T. X., Paques, K., Oliveira, J., Chiu, Y. D., Duckworth, M., Khialani, B., Kyranis, S., Bennett, M. R., West, N. E. J., & Hoole, S. P. (2024). Evaluation of microcirculatory protection in percutaneous revascularisation: A stent implantation technique and device comparison. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31155
Vancouver
Aetesam-Ur-Rahman M, Zhao TX, Paques K, Oliveira J, Chiu YD, Duckworth M, et al. Evaluation of microcirculatory protection in percutaneous revascularisation: A stent implantation technique and device comparison. Catheterization and Cardiovascular Interventions. 2024. doi:10.1002/ccd.31155
BibTeX
@article{AetesamUrRahman2024,
author = {Aetesam-Ur-Rahman, Muhammad and Zhao, Tian X and Paques, Kitty and Oliveira, Joana and Chiu, Yi-Da and Duckworth, Melissa and Khialani, Bharat and Kyranis, Stephen and Bennett, Martin R and West, Nick E J and Hoole, Stephen P},
title = {Evaluation of microcirculatory protection in percutaneous revascularisation: A stent implantation technique and device comparison},
journal = {Catheterization and Cardiovascular Interventions},
year = {2024},
doi = {10.1002/ccd.31155}
}
EUVIMED