EUVIMEDEUVIMED entdecken
Fallbericht

Extravascular protrusion of the Alterra adaptive prestent identified on surveillance computed tomography imaging

Matthew J Gillespie, Nicola Maschietto iD, Jamil A Aboulhosn, David T Balzer, Athar M Qureshi iD, Doff B McElhinney iD

Catheterization and Cardiovascular Interventions · 2024

DOI: 10.1002/ccd.31147 ↗

Kurz gefasst

Fallbericht in der Zeitschrift Catheterization and Cardiovascular Interventions, 2024. Verfasst von Matthew J Gillespie und 5 weiteren Autor:innen. Beteiligte Einrichtung: Division of Pediatric Cardiology, The Children's Hospital of Philadelphia, University of Pennsylvania School of Medicine, Philadelphia, Pennsylvania,…. Hauptthemen laut MEDLINE: Heart Valve Prosthesis, Heart Valve Prosthesis Implantation, Predictive Value of Tests, Prosthesis Design. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 10 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The Alterra adaptive prestent is a novel self‐expanding device designed to provide a landing zone for the 29 mm SAPIEN 3 valve to treat pulmonary regurgitation in patients with a right ventricular outflow tract that is too large for a balloon expandable valve alone. The mechanism of fixation for the Alterra prestent is radial force from the self‐expanding stent frame, combined with a unique set of flared “tines” that protrude from both ends of the stent. Aims, Methods, and Results In this report, we describe 6 patients who underwent uncomplicated transcatheter pulmonary valve replacement with an Alterra adaptive prestent and SAPIEN 3 valve and had surveillance chest computed tomography (CT) scans performed 1 day to 21 months after implant. In each patient, the CT scan demonstrated extravascular extension of a portion of the Alterra prestent, without clinical sequelae, but with extension into the ascending aorta in 1 patient and contact with the ascending aorta, left pulmonary vein, or left atrial appendage in 3 others.

Conclusions

Surveillance CT imaging shows that the Alterra prestent can perforate the pulmonary artery and/or right ventricle. Although no sequelae were seen in these patients, prestent perforation has the potential to be clinically important. Implanters should be aware of this finding and its potential implications. As experience with the Alterra prestent grows, it will be important to further define the risk factors, incidence, and implications of this phenomenon.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Heart Valve ProsthesisHeart Valve Prosthesis ImplantationPredictive Value of TestsProsthesis DesignAdolescentAdultHumansYoung AdultCardiac CatheterizationComputed Tomography AngiographyPulmonary ArteryPulmonary ValvePulmonary Valve InsufficiencyTime FactorsTomography, X-Ray ComputedTreatment Outcome

Stichwörter der Autor:innen

percutaneous valvepulmonary circulationpulmonary regurgitationpulmonary valvetetralogy of fallot

Autor:innen und Institutionen

  • Division of Pediatric Cardiology, The Children's Hospital of Philadelphia, University of Pennsylvania School of Medicine, Philadelphia, Pennsylvania, USA.
  • Department of Cardiology, Boston Children's Hospital, Boston, Massachusetts, USA.
  • Division of Pediatric Cardiology, Mattel Children's Hospital at UCLA, Ahmanson/UCLA Adult Congenital Heart Disease Center, Los Angeles, California, USA.
  • Department of Pediatrics, Washington University School of Medicine, St. Louis, Missouri, USA.
  • Division of Cardiology, Texas Children's Hospital and Department of Pediatrics, Baylor College of Medicine, Houston, Texas, USA.
  • Departments of Cardiothoracic Surgery and Pediatrics (Cardiology), Stanford University School of Medicine, Palo Alto, California, USA.

ORCID-Profile: Nicola Maschietto, Athar M Qureshi, Doff B McElhinney

Zitieren

APA

Gillespie, M. J., Maschietto, N., Aboulhosn, J. A., Balzer, D. T., Qureshi, A. M., & McElhinney, D. B. (2024). Extravascular protrusion of the Alterra adaptive prestent identified on surveillance computed tomography imaging. Catheterization and Cardiovascular Interventions. https://doi.org/10.1002/ccd.31147

Vancouver

Gillespie MJ, Maschietto N, Aboulhosn JA, Balzer DT, Qureshi AM, McElhinney DB. Extravascular protrusion of the Alterra adaptive prestent identified on surveillance computed tomography imaging. Catheterization and Cardiovascular Interventions. 2024. doi:10.1002/ccd.31147

BibTeX
@article{Gillespie2024,
  author = {Gillespie, Matthew J and Maschietto, Nicola and Aboulhosn, Jamil A and Balzer, David T and Qureshi, Athar M and McElhinney, Doff B},
  title = {Extravascular protrusion of the Alterra adaptive prestent identified on surveillance computed tomography imaging},
  journal = {Catheterization and Cardiovascular Interventions},
  year = {2024},
  doi = {10.1002/ccd.31147}
}
Kostenlos starten →