Validity of an ultrasound device to measure bone mineral density

Jessica L Puranda iD, Chris M Edwards, Vinicius M R Weber, Mohamed Aboudlal, Kevin Semeniuk, Kristi B Adamo

Clinical Anatomy · 2025

DOI: 10.1002/ca.24187 ↗

At a glance

Validierungsstudie in Clinical Anatomy, 2025. Written by Jessica L Puranda and 5 co-authors. Institution: School of Human Kinetics, Faculty of Health Sciences, University of Ottawa, Ottawa, Canada. Major MEDLINE topics: Bone Density, Ultrasonography, Absorptiometry, Photon. 26 references. Funded by: CIHR, Coordenação de Aperfeiçoamento de Pessoal de Nível Superior, Canadian Institutes of Health Research.

AbstraktSubjectsBibliografické údajeCiteRelated studies

Abstract (original language)

This study aims to examine the validity and reliability of the UltraScan650™, a portable ultrasound device, used to measure BMD at the 1/3rd radius position. Fifty‐two female first responders and healthcare providers were assessed using DXA (forearm, femur, lumbar, and total body) and the UltraScan650™. Fat and lean mass were also assessed using the DXA. Pearson correlations, Bland–Altman plots, t ‐tests, and linear regressions were used to assess validity. Intra‐class correlation (ICC) coefficients were used to assess reliability. Inter‐rater reliability and repeatability were good (ICC = 0.896 [0.818; 0.942], p < 0.001) and excellent (ICC = 0.917 [0.785; 0.989], p < 0.001), respectively. BMD as measured by the UltraScan650™ was weakly correlated to the DXA ( r = 0.382 [0.121; 0.593], p = 0.0052). Bland–Altman plots revealed that the UltraScan650™ underestimated BMD (−0.0569 g/cm 2 ), this was confirmed with a significant paired t ‐test ( p < 0.001). A linear regression was performed (0.4744 × UltraScan650™ + 0.4170) to provide more information as to the issue of agreement. Bland–Altman plots revealed a negligible bias, supported by a paired t ‐test ( p = 0.9978). Pearson's correlation revealed a significant relationship ( r = −0.771 [−0.862; −0.631], p < 0.0001) between adjusted UltraScan650™—DXA and the average of the two scans (i.e., adjusted UltraScan650™ and DXA), suggesting a proportional constant error and proportional constant variability in measurements of BMD from the UltraScan650™. The UltraScan650™ is not a valid alternative to DXA for diagnostic purposes; however, the UltraScan650™ could be used as a screening tool in the clinical and research setting given the linear transformation is employed.

Subjects

MeSH terms (MEDLINE)

Bone DensityUltrasonographyAbsorptiometry, PhotonHumansFemaleReproducibility of ResultsMiddle AgedAgedAdultRadius

Author keywords

bone mineral densitydiagnosticreliabilityultrasoundvalidation

Authors and institutions

  • School of Human Kinetics, Faculty of Health Sciences, University of Ottawa, Ottawa, Canada.
  • Laboratory of Experimental and Applied Physiology to Physical Activity, Midwest State University of Paraná, Guarapuava, Paraná, Brazil.

ORCID profiles: Jessica L Puranda

Funding

  • CIHR
  • Coordenação de Aperfeiçoamento de Pessoal de Nível Superior
  • Canadian Institutes of Health Research

Cite

APA

Puranda, J. L., Edwards, C. M., Weber, V. M. R., Aboudlal, M., Semeniuk, K., & Adamo, K. B. (2025). Validity of an ultrasound device to measure bone mineral density. Clinical Anatomy. https://doi.org/10.1002/ca.24187

Vancouver

Puranda JL, Edwards CM, Weber VMR, Aboudlal M, Semeniuk K, Adamo KB. Validity of an ultrasound device to measure bone mineral density. Clinical Anatomy. 2025. doi:10.1002/ca.24187

BibTeX
@article{Puranda2025,
  author = {Puranda, Jessica L and Edwards, Chris M and Weber, Vinicius M R and Aboudlal, Mohamed and Semeniuk, Kevin and Adamo, Kristi B},
  title = {Validity of an ultrasound device to measure bone mineral density},
  journal = {Clinical Anatomy},
  year = {2025},
  doi = {10.1002/ca.24187}
}
Start for free →