Validity of an ultrasound device to measure bone mineral density

Jessica L Puranda iD, Chris M Edwards, Vinicius M R Weber, Mohamed Aboudlal, Kevin Semeniuk, Kristi B Adamo

Clinical Anatomy · 2025

DOI: 10.1002/ca.24187 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Clinical Anatomy, 2025. Verfasst von Jessica L Puranda und 5 weiteren Autor:innen. Beteiligte Einrichtung: School of Human Kinetics, Faculty of Health Sciences, University of Ottawa, Ottawa, Canada. Hauptthemen laut MEDLINE: Bone Density, Ultrasonography, Absorptiometry, Photon. Die Arbeit nennt 26 Literaturangaben. Gefördert von: CIHR, Coordenação de Aperfeiçoamento de Pessoal de Nível Superior, Canadian Institutes of Health Research.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study aims to examine the validity and reliability of the UltraScan650™, a portable ultrasound device, used to measure BMD at the 1/3rd radius position. Fifty‐two female first responders and healthcare providers were assessed using DXA (forearm, femur, lumbar, and total body) and the UltraScan650™. Fat and lean mass were also assessed using the DXA. Pearson correlations, Bland–Altman plots, t ‐tests, and linear regressions were used to assess validity. Intra‐class correlation (ICC) coefficients were used to assess reliability. Inter‐rater reliability and repeatability were good (ICC = 0.896 [0.818; 0.942], p < 0.001) and excellent (ICC = 0.917 [0.785; 0.989], p < 0.001), respectively. BMD as measured by the UltraScan650™ was weakly correlated to the DXA ( r = 0.382 [0.121; 0.593], p = 0.0052). Bland–Altman plots revealed that the UltraScan650™ underestimated BMD (−0.0569 g/cm 2 ), this was confirmed with a significant paired t ‐test ( p < 0.001). A linear regression was performed (0.4744 × UltraScan650™ + 0.4170) to provide more information as to the issue of agreement. Bland–Altman plots revealed a negligible bias, supported by a paired t ‐test ( p = 0.9978). Pearson's correlation revealed a significant relationship ( r = −0.771 [−0.862; −0.631], p < 0.0001) between adjusted UltraScan650™—DXA and the average of the two scans (i.e., adjusted UltraScan650™ and DXA), suggesting a proportional constant error and proportional constant variability in measurements of BMD from the UltraScan650™. The UltraScan650™ is not a valid alternative to DXA for diagnostic purposes; however, the UltraScan650™ could be used as a screening tool in the clinical and research setting given the linear transformation is employed.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Bone DensityUltrasonographyAbsorptiometry, PhotonHumansFemaleReproducibility of ResultsMiddle AgedAgedAdultRadius

Stichwörter der Autor:innen

bone mineral densitydiagnosticreliabilityultrasoundvalidation

Autor:innen und Institutionen

  • School of Human Kinetics, Faculty of Health Sciences, University of Ottawa, Ottawa, Canada.
  • Laboratory of Experimental and Applied Physiology to Physical Activity, Midwest State University of Paraná, Guarapuava, Paraná, Brazil.

ORCID-Profile: Jessica L Puranda

Förderung

  • CIHR
  • Coordenação de Aperfeiçoamento de Pessoal de Nível Superior
  • Canadian Institutes of Health Research

Zitieren

APA

Puranda, J. L., Edwards, C. M., Weber, V. M. R., Aboudlal, M., Semeniuk, K., & Adamo, K. B. (2025). Validity of an ultrasound device to measure bone mineral density. Clinical Anatomy. https://doi.org/10.1002/ca.24187

Vancouver

Puranda JL, Edwards CM, Weber VMR, Aboudlal M, Semeniuk K, Adamo KB. Validity of an ultrasound device to measure bone mineral density. Clinical Anatomy. 2025. doi:10.1002/ca.24187

BibTeX
@article{Puranda2025,
  author = {Puranda, Jessica L and Edwards, Chris M and Weber, Vinicius M R and Aboudlal, Mohamed and Semeniuk, Kevin and Adamo, Kristi B},
  title = {Validity of an ultrasound device to measure bone mineral density},
  journal = {Clinical Anatomy},
  year = {2025},
  doi = {10.1002/ca.24187}
}
Kostenlos starten →