EUVIMEDEUVIMED entdecken
Crossref

An ultrahigh‐performance liquid chromatography–tandem mass spectrometry method for quantification of methotrexate and 7‐hydroxy‐methotrexate and application for therapeutic drug monitoring in patients with central nervous system lymphoma

Jiping Huo iD, Bo Zhang, Dongjie Zhang, Bin Zhu, Zhigang Zhao iD, Shenghui Mei iD

Biomedical Chromatography · 2024

DOI: 10.1002/bmc.5789 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biomedical Chromatography, 2024. Verfasst von Jiping Huo und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Pharmacy, Beijing Tiantan Hospital, Capital Medical University, Beijing, P. R. China. Hauptthemen laut MEDLINE: Methotrexate, Lymphoma. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 32 Literaturangaben. Gefördert von: Beijing Municipal Administration of Hospitals Incubating Program, Training Fund for Open Projects at Clinical Institutes and Departments of Capital Medical University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A method using ultrahigh‐performance liquid chromatography–tandem mass spectrometry was developed, validated, and applied to simultaneously determine plasma methotrexate (MTX) and 7‐hydroxy‐methotrexate (7‐OH‐MTX) in 117 patients with central nervous system (CNS) lymphoma. The ion transitions utilized were m/z 455.2 > 308.2 for MTX and m/z 471.2 > 324.1 for 7‐OH‐MTX. Samples were prepared through protein precipitation using methanol. Chromatographic separation was achieved within 3.0 min on a CMS9030 column (Ruixi, 2.1 × 50 mm, 3 μm) through a gradient elution of methanol and a 10% ammonium acetate solution at a flow rate of 0.4 mL/min. The method demonstrated linearity in the concentration range of 0.05–10 μM for MTX and 0.25–50 μM for 7‐OH‐MTX. The intra‐ and inter‐day inaccuracy ranged from –7.38% to 7.83%, and the imprecision was less than 6.00% for both analytes. The recovery and matrix effect normalized by the internal standard (MTX‐D 3 ) remained consistent. Both analytes remained stable under nine different storage conditions. In patients with CNS lymphoma, MTX levels at 12 h and 7‐OH‐MTX levels at 12, 36, and 60 h after dosing in individuals with impaired renal function were significantly higher compared with those with normal renal function. 7‐OH‐MTX could potentially serve as a superior indicator for nephrotoxicity compared with MTX.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

MethotrexateLymphomaHumansTandem Mass SpectrometryDrug MonitoringMethanolChromatography, High Pressure LiquidCentral Nervous System

Stichwörter der Autor:innen

7-hydroxy-methotrexateUHPLC-MS/MSmethod development and validationmethotrexaterenal impairment

Substanzen

MethotrexateMethanol

Autor:innen und Institutionen

  • Department of Pharmacy, Beijing Tiantan Hospital, Capital Medical University, Beijing, P. R. China.
  • Department of Clinical Pharmacology, College of Pharmaceutical Sciences, Capital Medical University, Beijing, Beijing, P. R. China.
  • Department of Pharmacy, Beijing Health Vocational College, Beijing, China.

ORCID-Profile: Jiping Huo, Zhigang Zhao, Shenghui Mei

Förderung

  • Beijing Municipal Administration of Hospitals Incubating Program
  • Training Fund for Open Projects at Clinical Institutes and Departments of Capital Medical University

Zitieren

APA

Huo, J., Zhang, B., Zhang, D., Zhu, B., Zhao, Z., & Mei, S. (2024). An ultrahigh‐performance liquid chromatography–tandem mass spectrometry method for quantification of methotrexate and 7‐hydroxy‐methotrexate and application for therapeutic drug monitoring in patients with central nervous system lymphoma. Biomedical Chromatography. https://doi.org/10.1002/bmc.5789

Vancouver

Huo J, Zhang B, Zhang D, Zhu B, Zhao Z, Mei S. An ultrahigh‐performance liquid chromatography–tandem mass spectrometry method for quantification of methotrexate and 7‐hydroxy‐methotrexate and application for therapeutic drug monitoring in patients with central nervous system lymphoma. Biomedical Chromatography. 2024. doi:10.1002/bmc.5789

BibTeX
@article{Huo2024,
  author = {Huo, Jiping and Zhang, Bo and Zhang, Dongjie and Zhu, Bin and Zhao, Zhigang and Mei, Shenghui},
  title = {An ultrahigh‐performance liquid chromatography–tandem mass spectrometry method for quantification of methotrexate and 7‐hydroxy‐methotrexate and application for therapeutic drug monitoring in patients with central nervous system lymphoma},
  journal = {Biomedical Chromatography},
  year = {2024},
  doi = {10.1002/bmc.5789}
}
Kostenlos starten →