EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC 4.0

Wireless blood pressure sensor implantation in sheep: A detailed technique

David Santer iD, Paul Ecker, Judith Radloff, Katharina Heissl, Smruti Mirchandani, Sabine Leitner, Bruno Mora iD, Martin Czerny, Roman Gottardi, Michael Moore, Bruno K Podesser

Animal Models and Experimental Medicine · 2026-03 · 9 (3) · 630-638

DOI: 10.1002/ame2.70159 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Animal Models and Experimental Medicine, online erschienen am 06.03.2026. Verfasst von David Santer und 10 weiteren Autor:innen. Beteiligte Einrichtung: Center for Biomedical Research and Translational Surgery, Medical University of Vienna, Vienna, Austria. Hauptthemen laut MEDLINE: Blood Pressure, Telemetry, Blood Pressure Monitors, Wireless Technology. Die Arbeit nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Wirtschaftsagentur Wien (WAW, Mariahilfer Straße 20, 1070 Wien), Program Title: CallLife Sciences 2021, Wirtschaftsagentur Wien.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Biomedical research and preclinical testing in large animal models are essential for developing and evaluating new medical devices. This ovine study presents a stepwise implantation protocol and initial experiences with a wireless blood pressure implant sensor, enabling continuous telemetric monitoring in sheep. Adult female Jezersko‐Solčava sheep underwent implantation of a wireless blood pressure sensor into the ascending aorta via the right carotid artery. The body of the device can was positioned subcutaneously in the lower neck region. After skin closure, a computed tomography scan was performed to verify sensor tip location. If malposition was detected, surgical revision was undertaken immediately. Wireless blood pressure monitoring was conducted over a 24 h period in a custom‐designed sheep barn. Twelve adult female sheep successfully underwent the implantation procedure, with no major complications reported. Sensor catheter revision was required in three animals (25%). The mean duration of the procedure (from incision to skin closure) was 36 ± 10 min. During the 24 h monitoring period, the average systolic and diastolic blood pressures were 93 ± 6 and 65 ± 5 mmHg, respectively. Large animal models, particularly sheep, are indispensable in cardiovascular device research due to their anatomical and physiological similarities to humans. This study demonstrates a rapid, minimally invasive protocol for wireless blood pressure sensor implantation in sheep, offering a valuable platform for chronic telemetric monitoring in the development of cardiovascular medical devices.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood PressureTelemetryBlood Pressure MonitorsWireless TechnologyBlood Pressure DeterminationAnimalsFemaleSheep

Stichwörter der Autor:innen

biomedical researchblood pressure monitoringlarge animal modelsheep

Autor:innen und Institutionen

  • Center for Biomedical Research and Translational Surgery, Medical University of Vienna, Vienna, Austria.
  • Medical Faculty of the University of Basel, Basel, Switzerland.
  • Ascense Medical GmbH, Vienna, Austria.
  • Department of Thoracic and Cardiovascular Anesthesia and Intensive Care, General Hospital Vienna, Medical University of Vienna, Vienna, Austria.
  • Department of Cardiovascular Surgery, University Heart Center Freiburg-Bad Krozingen, Freiburg, Germany.
  • Faculty of Medicine, Albert Ludwigs University Freiburg, Freiburg, Germany.

ORCID-Profile: David Santer, Bruno Mora

Förderung

  • Wirtschaftsagentur Wien (WAW, Mariahilfer Straße 20, 1070 Wien), Program Title: CallLife Sciences 2021
  • Wirtschaftsagentur Wien

Zitieren

APA

Santer, D., Ecker, P., Radloff, J., Heissl, K., Mirchandani, S., Leitner, S., Mora, B., Czerny, M., Gottardi, R., Moore, M., & Podesser, B. K. (2026). Wireless blood pressure sensor implantation in sheep: A detailed technique. Animal Models and Experimental Medicine, 9(3), 630-638. https://doi.org/10.1002/ame2.70159

Vancouver

Santer D, Ecker P, Radloff J, Heissl K, Mirchandani S, Leitner S, et al. Wireless blood pressure sensor implantation in sheep: A detailed technique. Animal Models and Experimental Medicine. 2026;9(3):630-638. doi:10.1002/ame2.70159

BibTeX
@article{Santer2026,
  author = {Santer, David and Ecker, Paul and Radloff, Judith and Heissl, Katharina and Mirchandani, Smruti and Leitner, Sabine and Mora, Bruno and Czerny, Martin and Gottardi, Roman and Moore, Michael and Podesser, Bruno K},
  title = {Wireless blood pressure sensor implantation in sheep: A detailed technique},
  journal = {Animal Models and Experimental Medicine},
  year = {2026},
  volume = {9},
  number = {3},
  pages = {630--638},
  doi = {10.1002/ame2.70159}
}
Kostenlos starten →