Fachartikel in der Zeitschrift Advanced Healthcare Materials, 2025. Verfasst von Zhengdi Wang und 7 weiteren Autor:innen. Beteiligte Einrichtung: Experimental Center of Advanced Materials, School of Materials Science & Engineering, Beijing Institute of Technology, Beijing, China. Hauptthemen laut MEDLINE: Biosensing Techniques, Wearable Electronic Devices, Glucose, Sweat. Die Arbeit nennt 30 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Beijing Natural Science Foundation, Beijing Institute of Technology Research Fund Program for Young Scholars.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Wearable electrochemical glucose sensors face critical challenges in balancing enzyme stability, electron transfer efficiency, and mechanical durability. In this study, we present a flexible glucose‐sensing patch based on a boron carbide (B 4 C) nanoskeleton grown directly on activated cotton textiles (ACT) via programmable vapor–liquid–solid (VLS) synthesis. By precisely tuning nickel catalyst size, interparticle spacing, and B:Ni molar ratio, we engineered nest‐like 3D B 4 C nanowire networks that preserve the ACT substrate's inherent flexibility and hierarchical porosity. This architecture ensures continuous electron conduction and supports hydrogen‐bond‐driven immobilization of glucose oxidase (GOx) through in situ‐generated ─NH 2 /─OH groups, eliminating the need for additional chemical modifications. The resulting B 4 C‐ACT@GOx electrode exhibits high sensitivity (36.288 µA mM −1 cm −2 ) within the physiological sweat glucose ranges (5 µM–1 m M ), an ultrafast response time of 0.1 s, and longterm stability over 4 weeks. Integrated into a wireless patch, the device enables real‐time glucose monitoring in human sweat. This work bridges nanoscale material engineering and wearable biosensor functionality, providing a scalable platform for personalized healthcare applications.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Biosensing TechniquesWearable Electronic DevicesGlucoseSweatBoron CompoundsHumansGlucose OxidaseEnzymes, ImmobilizedNanowiresElectrodesElectrochemical Techniques

Stichwörter der Autor:innen

3D nanowire networksboron carbide nanoskeletonhydrogen‐bonded GOx immobilizationprogrammable VLS synthesissweat glucose monitoring

Substanzen

GlucoseGlucose OxidaseEnzymes, ImmobilizedBoron Compounds

Autor:innen und Institutionen

  • Experimental Center of Advanced Materials, School of Materials Science & Engineering, Beijing Institute of Technology, Beijing, China.
  • School of Automation, Beijing Institute of Technology, Beijing, China.

ORCID-Profile: Minmin Liang

Förderung

  • National Natural Science Foundation of China
  • Beijing Natural Science Foundation
  • Beijing Institute of Technology Research Fund Program for Young Scholars
  • Natural Science Foundation of Beijing Municipality

Zitieren

APA

Wang, Z., Wang, X., Wen, H., He, Z., Guo, Z., Yin, S., Song, N., & Liang, M. (2025). Boron Carbide Nanoskeleton‐Engineered Wearable Biosensor for Real‐Time Sweat Glucose Monitoring. Advanced Healthcare Materials. https://doi.org/10.1002/adhm.202503624

Vancouver

Wang Z, Wang X, Wen H, He Z, Guo Z, Yin S, et al. Boron Carbide Nanoskeleton‐Engineered Wearable Biosensor for Real‐Time Sweat Glucose Monitoring. Advanced Healthcare Materials. 2025. doi:10.1002/adhm.202503624

BibTeX
@article{Wang2025,
  author = {Wang, Zhengdi and Wang, Xiaoyan and Wen, Hailong and He, Zhu and Guo, Zhanjun and Yin, Sijie and Song, Ningning and Liang, Minmin},
  title = {Boron Carbide Nanoskeleton‐Engineered Wearable Biosensor for Real‐Time Sweat Glucose Monitoring},
  journal = {Advanced Healthcare Materials},
  year = {2025},
  doi = {10.1002/adhm.202503624}
}
Kostenlos starten →