Are two readings as good as three for monthly linear accelerator output quality assurance?
Journal of Applied Clinical Medical Physics · 2026-12 · 27 (12)
DOI: 10.1002/acm2.70912 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 22.11.2026. Verfasst von Ethan D. Stolen und 5 weiteren Autor:innen. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
During monthly linear accelerator (linac) output quality assurance (QA), it is common practice to acquire three consecutive ionization chamber readings per beam energy and average them. Whether two readings carry the same QA decisions in practice has not been directly quantified against historical output data.
Purpose
This study evaluates whether reducing the number of readings per monthly output measurement from three to two preserves QA decision accuracy against the TG‐142 ± 2% action threshold. Within‐session variation was additionally evaluated against the ± 0.5% between‐reading criterion of TG‐198.
Methods
A retrospective analysis was performed on three years of monthly output QA data from four linear accelerators at a single institution, across eight beam energies (6, 9, 12, and 16 MeV electrons; 6, 6 FFF, 10, and 15 MV photons), with each check comprising three consecutive readings per energy. The analysis combined four components: (i) a paired‐difference comparison of per‐session means from two‐reading versus three‐reading protocols, summarized with mean differences, percent differences, and Cohen's d; (ii) an evaluation of within‐session reading variation against the TG‐198 criterion; (iii) a scenario analysis of false‐positive and false‐negative classifications against the TG‐142 action threshold, with the three‐reading decision taken as ground truth; and (iv) generalized linear mixed‐effects modeling of individual readings, with random intercepts for machine and measurement session, to test for systematic reading‐order shifts.
Results
Across all eight energies, the absolute mean difference between two‐ and three‐reading protocols was at most 0.006 nC, corresponding to less than 0.04% of the corrected output. The two protocols produced concordant pass/fail decisions in 961 of 963 recorded sessions (99.1–100% per energy). The two‐reading protocol produced no false negatives: the two sessions above the action threshold were detected by both protocols, and the only disagreements were two isolated false alarms (one at 6 MeV, one at 6 MV). Within‐session reading variation satisfied the TG‐198 ± 0.5% criterion in 91.9–100% of sessions per energy, and restricting attention to the first two readings never increased the fraction of sessions exceeding that criterion. Mixed‐effects modeling identified a small but statistically detectable reading‐order shift: relative to the first reading, the second and third photon readings were lower by approximately 0.05% and 0.12%, and the corresponding electron shifts were near 0.05%. In all cases the absolute magnitude of these shifts was below 0.18%, well within the ± 0.5% between‐reading variation limit recommended by TG‐198.
Conclusions
For the eight beam energies studied, monthly output QA decisions based on two readings were consistent with those based on three readings, with no false negatives and a reading‐order effect well within the TG‐198 ± 0.5% limit. Low‐energy electron beams (non‐normal repeatability, the largest single outlier) and 10 MV (most frequent exceedance of the 0.5% criterion) warrant closer attention. These results indicate that two‐reading protocols can maintain QA decision accuracy while reducing measurement time.
Zitieren
APA
Stolen, E. D., Schulz, J. B., Stolen, E. D., Paek, A., Kim, S., & Sohn, J. J. (2026). Are two readings as good as three for monthly linear accelerator output quality assurance?. Journal of Applied Clinical Medical Physics, 27(12). https://doi.org/10.1002/acm2.70912
Vancouver
Stolen ED, Schulz JB, Stolen ED, Paek A, Kim S, Sohn JJ. Are two readings as good as three for monthly linear accelerator output quality assurance?. Journal of Applied Clinical Medical Physics. 2026;27(12). doi:10.1002/acm2.70912
BibTeX
@article{Stolen2026,
author = {Stolen, Ethan D. and Schulz, Joseph B. and Stolen, Eric D. and Paek, Ava and Kim, Siyong and Sohn, James J.},
title = {Are two readings as good as three for monthly linear accelerator output quality assurance?},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {12},
doi = {10.1002/acm2.70912}
}
EUVIMED