Implementation of a 0.55T MR simulator for intracranial stereotactic radiosurgery

Joshua Kim, Robert Rusnac, Amanda Dicarlo, Anthony Doemer, Ralph Ruiz, M. Salim Siddiqui, M. Shah, Eric A. Mellon, Benjamin Movsas, Kundan Thind

Journal of Applied Clinical Medical Physics · 2026-12 · 27 (12)

DOI: 10.1002/acm2.70909 ↗

At a glance

Journal article in Journal of Applied Clinical Medical Physics, published online on 2026-11-19. Written by Joshua Kim and 9 co-authors. 15 references. The full text is freely available (open access, CC BY 4.0).

AbstractBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Low‐field MRI has several advantages for radiation therapy simulation. The reduced field strength lowers susceptibility artifacts near metallic implants, decreases intrinsic geometric distortion, and improves safety through lower SAR, reduced projectile risk, and fewer device interactions. MRI is the imaging foundation for SRS target delineation, and initial low‐field MRI feasibility for brain radiotherapy planning has been demonstrated in small cohorts. However, SRS clinical implementation has not been reported. We describe the implementation of a 0.55T MR simulator and characterize lesion conspicuity across 102 lesions in 13 patients with multiple brain metastases, drawn from 211 consecutive SRS patients treated to 511 lesions over 12 months.

Purpose

To describe the clinical implementation of a 0.55T MR simulator (MAGNETOM Free.Max RT Edition, Siemens Healthineers) as a dedicated radiation therapy simulation system, following a systematic framework of acceptance testing per AAPM TG‐284, longitudinal quality assurance, and SRS MRI protocol optimization.

Methods

Acceptance testing and commissioning followed AAPM TG‐284 recommendations. Testing included the use of vendor‐provided homogeneous phantoms, the ACR large phantom, and a commercially available large‐field geometric distortion phantom. Longitudinal quality assurance was performed over a 12‐month period using the ACR large phantom. Measured parameters included geometric accuracy, percent image uniformity (PIU), percent signal ghosting, low contrast detectability (LCD), and SNR. Contrast to noise ratio (CNR) was calculated for the lesions of a subset of SRS patients having 4 or greater metastatic lesions to evaluate the conspicuity of small SRS lesions in this challenging environment.

Results

All acceptance testing met TG‐284 recommendations. Central frequency and transmitter gain were stable with changes of less than 0.01% and 0.3%, respectively, over the first month of testing. Geometric distortion after vendor 3D corrections measured less than 1 mm within 10 cm and less than 2 mm within 20 cm of isocenter. Monthly phantom testing over 12 months confirmed geometric stability within established tolerances. A brain imaging protocol achieved 1 mm slice thickness on 3D T1‐weighted acquisitions consistent with ISRS guidelines, with total scan time under 35 minutes. To date, 211 intracranial SRS patients have been scanned and treated to a total of 511 lesions with a median volume of 0.58cc [0.004 – 304.48 cc]. The median CNR for the 102 lesions in the set of 13 patients with 4 or greater metastatic lesions was calculated as 6.81 [1.7–16.9] with only two lesions having a CNR < 2.

Conclusions

These findings support the feasibility of 0.55T MRI as a dedicated RT simulator capable of meeting the imaging demands of intracranial SRS. Formal comparative studies against higher field systems are needed before low‐field imaging can be considered a standalone replacement for SRS planning.

Cite

APA

Kim, J., Rusnac, R., Dicarlo, A., Doemer, A., Ruiz, R., Siddiqui, M. S., Shah, M., Mellon, E. A., Movsas, B., & Thind, K. (2026). Implementation of a 0.55T MR simulator for intracranial stereotactic radiosurgery. Journal of Applied Clinical Medical Physics, 27(12). https://doi.org/10.1002/acm2.70909

Vancouver

Kim J, Rusnac R, Dicarlo A, Doemer A, Ruiz R, Siddiqui MS, et al. Implementation of a 0.55T MR simulator for intracranial stereotactic radiosurgery. Journal of Applied Clinical Medical Physics. 2026;27(12). doi:10.1002/acm2.70909

BibTeX
@article{Kim2026,
  author = {Kim, Joshua and Rusnac, Robert and Dicarlo, Amanda and Doemer, Anthony and Ruiz, Ralph and Siddiqui, M. Salim and Shah, M. and Mellon, Eric A. and Movsas, Benjamin and Thind, Kundan},
  title = {Implementation of a 0.55T MR simulator for intracranial stereotactic radiosurgery},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {12},
  doi = {10.1002/acm2.70909}
}
Start for free →