EUVIMEDOpdag EUVIMED
Open access · CC BY 4.0

Comparison of visual and semiautomated analysis techniques for depth of visualization assessment in ultrasound quality assurance

Miia Nadhum, Anna Vuorenmaa, Ullamari Hakulinen, Antti Vehkaoja

Journal of Applied Clinical Medical Physics · 2026-12 · 27 (12)

DOI: 10.1002/acm2.70883 ↗

At a glance

Journal article in Journal of Applied Clinical Medical Physics, published online on 2026-11-20. Written by Miia Nadhum and 3 co-authors. 21 references. The full text is freely available (open access, CC BY 4.0).

AbstractBibliografiske oplysningerCiteRelated studies

Abstract (original language)

Background

Depth of visualization (DoV) measurements are recommended as part of routine ultrasound (US) quality assurance (QA) in international guidelines. Effective QA requires measurement and analysis procedures that are repeatable, reliable, and minimally operator‑dependent.

Purpose

This study compares three visual DoV analysis methods with semiautomated analysis tool UltraIQ to identify the approach with the highest repeatability and improved objectivity.

Methods

A total of 215 retrospective DoV measurements acquired between 2018–2022 were analyzed, including 40 transducer models from 14 US devices by six manufacturers. Visual DoV measurements were performed on the US system display and on a medical‐grade display using a Picture Archiving and Communication System (PACS) using the vertical distance group and the uniformity images obtained from QA Phantom. DoV was also evaluated using SNR‐based analysis in the UltraIQ software. Repeatability was evaluated using three repeated DoV measurements for 17 transducers, and reliability was assessed using the intraclass correlation coefficient (ICC).

Results

UltraIQ provided DoV values that were 5.0%–14.5% deeper than those obtained with visual methods. Visual DoV on the US device display was 0.7–2.0 cm shallower than UltraIQ and 0.1–0.9 cm shallower than visual DoV on PACS. Visual measurements using uniformity images on PACS were 0.4–1.1 cm shallower than UltraIQ but demonstrated the closest agreement with UltraIQ results. ICC analysis showed good to excellent reliability. For visually measured DoV, ICC values ranged 0.893–0.939, while UltraIQ provided an ICC of 0.882.

Conclusion

Semiautomated DoV analysis using UltraIQ is as repeatable and reliable as visual assessment while providing a more objective approach for US QA.

Cite

APA

Nadhum, M., Vuorenmaa, A., Hakulinen, U., & Vehkaoja, A. (2026). Comparison of visual and semiautomated analysis techniques for depth of visualization assessment in ultrasound quality assurance. Journal of Applied Clinical Medical Physics, 27(12). https://doi.org/10.1002/acm2.70883

Vancouver

Nadhum M, Vuorenmaa A, Hakulinen U, Vehkaoja A. Comparison of visual and semiautomated analysis techniques for depth of visualization assessment in ultrasound quality assurance. Journal of Applied Clinical Medical Physics. 2026;27(12). doi:10.1002/acm2.70883

BibTeX
@article{Nadhum2026,
  author = {Nadhum, Miia and Vuorenmaa, Anna and Hakulinen, Ullamari and Vehkaoja, Antti},
  title = {Comparison of visual and semiautomated analysis techniques for depth of visualization assessment in ultrasound quality assurance},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {12},
  doi = {10.1002/acm2.70883}
}
Start for free →