Retrospective log‐based quality assurance of real‐time MLC adaptation on a helical tomotherapy platform
Journal of Applied Clinical Medical Physics · 2026
DOI: 10.1002/acm2.70855 ↗At a glance
Journal article in Journal of Applied Clinical Medical Physics, 2026. Written by Scott B Crowe and 1 co-authors. Institution: Cancer Care Services, Royal Brisbane and Women's Hospital, Brisbane, Australia. Major MEDLINE topics: Radiotherapy, Intensity-Modulated, Quality Assurance, Health Care, Radiotherapy Planning, Computer-Assisted, Prostatic Neoplasms. 32 references. The full text is freely available (open access, CC BY 4.0).
Abstract (original language)
Background
Real‐time multileaf collimator (MLC) adaptation is available on some helical tomotherapy treatment units. However, these systems are not supplied with software that facilitates the retrospective review of adapted delivery data.
Purpose
This study used in‐house software to complete a retrospective review of 1,566 treatment fractions with the aim of verifying the suitability and accuracy of MLC adaptations provided by a real‐time target tracking system.
Methods
A data extractor tool was used to access files containing planned MLC sinogram, the telemetry treatment MLC sinogram, the telemetry delivery timing data, and the 3‐dimensional target offset data from each treatment fraction. These data were used to calculate target offsets in the beam's eye view so that expected adaptations could be evaluated and compared to the adaptations recorded by the MLC telemetry system. Further investigations were undertaken using dosimeter array measurements with a motion platform.
Results
While MLC telemetry matched expected adaptations for the vast majority of treatment fractions, erroneous MLC leaf motions lasting >1 s were identified in four treatment fractions distributed across one lung and two prostate patients (amounting to 0.3% of the total number of fractions examined), producing dose discrepancies of up to 0.3 Gy which equated to 0.6% of the total prescription dose from the affected lung treatment. Phantom measurements demonstrated that a patch provided by the vendor successfully reduced the dosimetric impact of these errors to a negligible level.
Conclusions
A retrospective review of 1,566 treatment fractions led to the identification of a rare error which was communicated to and subsequently patched worldwide by the vendor, exemplifying the importance of post‐treatment quality assurance of adaptive radiotherapy.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Cancer Care Services, Royal Brisbane and Women's Hospital, Brisbane, Australia.
- School of Electrical Engineering and Computer Science, University of Queensland, Brisbane, Australia.
- School of Chemistry and Physics, Queensland University of Technology, Brisbane, Australia.
Cite
APA
Crowe, S. B., & Kairn, T. (2026). Retrospective log‐based quality assurance of real‐time MLC adaptation on a helical tomotherapy platform. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70855
Vancouver
Crowe SB, Kairn T. Retrospective log‐based quality assurance of real‐time MLC adaptation on a helical tomotherapy platform. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70855
BibTeX
@article{Crowe2026,
author = {Crowe, Scott B and Kairn, Tanya},
title = {Retrospective log‐based quality assurance of real‐time MLC adaptation on a helical tomotherapy platform},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70855}
}
EUVIMED