Reproducibility and positioning sensitivity of CT beam width measurements using a pencil ionization chamber and radiopaque mask

Morgan Aire, Hunter C Meyer, Krystal Kirby

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70753 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Morgan Aire und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics and Astronomy, Louisiana State University, Baton Rouge, Louisiana, USA. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Phantoms, Imaging, Radiotherapy Planning, Computer-Assisted, Radiometry. Die Arbeit nennt 10 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To evaluate the reproducibility and precision of CT beam width measurements using a pencil ionization chamber and radiopaque mask and to assess the robustness of the technique against clinically realistic setup errors in the superior‐inferior (SI) and anterior‐posterior (AP) directions.

Methods

Beam width was measured on a GE Discovery RT590 CT scanner at three nominal collimations (10, 15, 20 mm), three tube potentials (80, 100, 120 kV), and three tube currents (50, 100, 150 mA). Three repeated exposures per setting were acquired using a Radcal Accu‐Gold pencil ionization chamber with a radiopaque mask to calculate beam width. Coefficients of variation and Levene's test were used to test precision. To assess the robustness to setup error, the chamber was offset from the isocenter at 0, 1, 3, and 5 mm in the SI direction at 20‐mm collimation across all kV/mA combinations and at 0 and 5 mm in the AP direction at 10‐ and 20‐mm collimation. A linear mixed‐effects model was used to test the sensitivity of beam width to SI setup errors across kV/mA combinations. Slopes of the overall deviance from the global mean vs offset were obtained for each combination with 95% confidence intervals, and pairwise slope differences were evaluated using a Tukey‐adjusted statistical test. Confidence intervals were used to assess AP setup errors at two different nominal collimations.

Results

Across all kV/mA combinations, mean measured radiation beam width was 13.04, 16.90, and 20.60 mm for nominal 10‐, 15‐, and 20‐mm collimations, with inter‐protocol coefficients of variation of 0.57%, 0.48%, and 0.25%, respectively. Intra‐protocol standard deviations across triplicate exposures were 0.04 to 0.09 mm. SI offset slopes from 0 – 3 mm ranged from −0.07 to −0.12 mm (per mm of offset) with overlapping 95% confidence intervals. There was no significant interaction with kV/mA ( p > 0.05). AP offsets of 5 mm produced beam‐width changes of 0.09 mm at 10 mm collimation and 0.15 mm at 20 mm collimation.

Conclusion

The pencil ionization chamber and radiopaque mask technique produced highly reproducible CT beam width estimates for potentials up to 120 kV and at or above 50 mA that were independent of tube potential, tube current, and robust to clinically plausible setup inconsistencies.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tomography, X-Ray ComputedPhantoms, ImagingRadiotherapy Planning, Computer-AssistedRadiometryPatient PositioningHumansReproducibility of ResultsRadiotherapy DosageRadiotherapy, Intensity-Modulated

Stichwörter der Autor:innen

beam widthcomputed tomographyionization chamberquality assurancereproducibility

Autor:innen und Institutionen

  • Department of Physics and Astronomy, Louisiana State University, Baton Rouge, Louisiana, USA.
  • Department of Radiation Oncology, Mary Bird Perkins Cancer Center, Baton Rouge, Louisiana, USA.

Zitieren

APA

Aire, M., Meyer, H. C., & Kirby, K. (2026). Reproducibility and positioning sensitivity of CT beam width measurements using a pencil ionization chamber and radiopaque mask. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70753

Vancouver

Aire M, Meyer HC, Kirby K. Reproducibility and positioning sensitivity of CT beam width measurements using a pencil ionization chamber and radiopaque mask. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70753

BibTeX
@article{Aire2026,
  author = {Aire, Morgan and Meyer, Hunter C and Kirby, Krystal},
  title = {Reproducibility and positioning sensitivity of CT beam width measurements using a pencil ionization chamber and radiopaque mask},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70753}
}
Kostenlos starten →