EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Impact of MRI slice thickness and fiducial marker orientation on geometric accuracy in multi‐echo GRE imaging for prostate radiotherapy planning

Aika Hayashi, Yasukata Takahashi, Naoko Sanuki, Yasufumi Yamashita, Yoshitaka Suzuki, Kanae Matsuura, Hiroe Muto

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70741 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Aika Hayashi und 6 weiteren Autor:innen. Beteiligte Einrichtung: Graduate School of Medical Science, Suzuka University of Medical Science, Suzuka, Mie, Japan. Hauptthemen laut MEDLINE: Prostatic Neoplasms, Fiducial Markers, Magnetic Resonance Imaging, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Precise target definition in prostate radiotherapy requires accurate computed tomography (CT)‐magnetic resonance imaging (MRI) image registration using gold fiducial markers (GFMs). While multi‐echo gradient recalled echo (ME GRE) sequences facilitate GFM identification, the quantitative impact of acquisition parameters on physical measurement accuracy remains insufficiently understood.

Purpose

This study evaluates how MRI slice thickness and marker orientation angle jointly affect GFM visualization and geometric accuracy to characterize the resulting localization uncertainties.

Methods

A kiwifruit‐based phantom (T1: 1603.3 ms; T2: 72.3 ms) that mimics the relaxation times of the human prostate was imaged with 3.0T MRI. Imaging was performed using a 3D ME GRE sequence with three slice thicknesses (1.0, 2.0, and 3.0 mm) and 12 orientation angles ranging from horizontal to 90°. Five multidisciplinary observers independently identified marker coordinates while blinded to the acquisition parameters. Subjective confidence scores were also recorded.

Results

Two‐way analysis of variance (ANOVA) revealed significant main effects for slice thickness ( F (2, 144) = 32.03, p < 0.001) and orientation angle ( F (11, 144) = 4.06, p < 0.001), with significant interaction between them ( F (22, 144) = 2.72, p < 0.001). Mean measurement errors with 95% confidence intervals (CI) were 0.96 mm (95% CI: 0.70–1.22) for 1.0 mm, 1.49 mm (95% CI: 1.23–1.75) for 2.0 mm, and 2.00 mm (95% CI: 1.74–2.26) for 3.0 mm slices. The peak error reached 3.10 mm under the combination of a 3.0 mm slice and 15° orientation. Subjective confidence scores showed a significant negative correlation with physical measurement errors ( r = −0.296, p < 0.001).

Conclusions

MRI slice thickness and marker orientation significantly impact GFM visualization accuracy. The results suggest that a 1.0 mm slice thickness offers a technical advantage for maintaining sub‐millimeter localization accuracy across various marker orientations and highly desirable for high‐precision radiotherapy planning.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Prostatic NeoplasmsFiducial MarkersMagnetic Resonance ImagingRadiotherapy Planning, Computer-AssistedPhantoms, ImagingImage Processing, Computer-AssistedHumansMaleRadiotherapy, Intensity-ModulatedRadiotherapy DosageTomography, X-Ray ComputedRadiotherapy, Image-Guided

Stichwörter der Autor:innen

Gold fiducial markersImage registrationMagnetic resonance imagingProstate cancer

Autor:innen und Institutionen

  • Graduate School of Medical Science, Suzuka University of Medical Science, Suzuka, Mie, Japan.
  • Department of Central Radiology, Yokkaichi Municipal Hospital, Yokkaichi, Mie, Japan.
  • Department of Radiology, Keio University School of Medicine, Shinjuku-ku, Tokyo, Japan.
  • Department of Radiology, Yokkaichi Municipal Hospital, Yokkaichi, Mie, Japan.
  • Department of Radiology, Mie University Hospital, Tsu, Mie, Japan.
  • Department of Radiological Technology, Faculty of Health Science, Suzuka University of Medical Science, Suzuka, Mie, Japan.

Zitieren

APA

Hayashi, A., Takahashi, Y., Sanuki, N., Yamashita, Y., Suzuki, Y., Matsuura, K., & Muto, H. (2026). Impact of MRI slice thickness and fiducial marker orientation on geometric accuracy in multi‐echo GRE imaging for prostate radiotherapy planning. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70741

Vancouver

Hayashi A, Takahashi Y, Sanuki N, Yamashita Y, Suzuki Y, Matsuura K, et al. Impact of MRI slice thickness and fiducial marker orientation on geometric accuracy in multi‐echo GRE imaging for prostate radiotherapy planning. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70741

BibTeX
@article{Hayashi2026,
  author = {Hayashi, Aika and Takahashi, Yasukata and Sanuki, Naoko and Yamashita, Yasufumi and Suzuki, Yoshitaka and Matsuura, Kanae and Muto, Hiroe},
  title = {Impact of MRI slice thickness and fiducial marker orientation on geometric accuracy in multi‐echo GRE imaging for prostate radiotherapy planning},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70741}
}
Kostenlos starten →