Potential of quantitative X‐ray imaging from photon‐counting CT: A novel analysis based on effective atomic number and physical density
Journal of Applied Clinical Medical Physics · 2026-07 · 27 (7)
DOI: 10.1002/acm2.70673 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 07.07.2026. Verfasst von Takashi Asahara und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Technology, Faculty of Health Sciences, Okayama University, Okayama, Japan. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Photons, Algorithms, Phantoms, Imaging. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: KAKENHI, Teraoka Scholarship Foundation, Cardiovascular Research Fund.
Abstract (Originalsprache)
Background
Conventional CT assessment of lung lesions is based mainly on morphological features. CT values represent relative X‐ray attenuation that does not directly reflect tissue composition and/or physical density. Quantitative imaging using photon‐counting CT (PC‐CT) has the potential to provide quantitative parameters, such as effective atomic number (Z eff ) and effective physical density (ρ eff ).
Purpose
The purpose is to develop and demonstrate the potential of an algorithm that generates both the Z eff and ρ eff images, derived directly from virtual monoenergetic images (VMIs) acquired with a clinical PC‐CT system.
Methods
The ρ eff image was generated by fitting a database of mass attenuation coefficients (μ/ρ) to the measured linear attenuation coefficients (μ), which were obtained from two VMIs at 70 and 100 keV. As an effective material, Z eff information was also determined and fed back into the ρ eff calculation. An in‐house low‐density phantom was scanned for quantitative evaluation of ρ eff . In addition, to demonstrate the applicability of our procedure to actual clinical imaging, representative chest PC‐CT images from patients were analyzed.
Results
In experiments using low‐density phantoms, the CT values did not necessarily correlate with Z eff values, but they showed a very good correlation with the ρ eff values. The ρ eff values calculated using our procedure correlated well with the correct values of ρ. A relative root mean square error of ρ eff calculation was 5.5%. Furthermore, we found that in pulmonary diagnosis, image contrast information from ρ eff makes it easier to identify and distinguish lesions compared to that from Z eff .
Conclusions
The proposed procedure directly generated Z eff and ρ eff images from PC‐CT data and enabled quantitative interpretation of CT value in terms of elemental composition and physical density. In this study, the feasibility of generating Z eff and ρ eff images was demonstrated.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiological Technology, Faculty of Health Sciences, Okayama University, Okayama, Japan.
- Graduate School of Medical Sciences, Kanazawa University, Kanazawa, Ishikawa, Japan.
- Department of Radiological Science, Faculty of Health Sciences, Junshin Gakuen University, Fukuoka, Japan.
- Division of Radiological Technology, Medical Support Department, Okayama University Hospital, Okayama, Japan.
- Department of Radiology, Medical Development Field, Okayama University, Okayama, Japan.
- Department of Radiology, Faculty of Medicine, Dentistry and Pharmaceutical Sciences, Okayama University, Okayama, Japan.
ORCID-Profile: Takashi Asahara, Hiroaki Hayashi
Förderung
- KAKENHI
- Teraoka Scholarship Foundation
- Cardiovascular Research Fund
- Kurata Grants from the Hitachi Global Foundation
- Grant for Young Researchers from the JGC-S (Nikki-Saneyoshi) Scholarship Foundation
- Suzuki Memorial Foundation
- Japan Society for the Promotion of Science
Zitieren
APA
Asahara, T., Nishigami, R., Kimoto, N., Mitani, M., Morimitsu, Y., Akagi, N., Higaki, F., Iguchi, T., Hiraki, T., & Hayashi, H. (2026). Potential of quantitative X‐ray imaging from photon‐counting CT: A novel analysis based on effective atomic number and physical density. Journal of Applied Clinical Medical Physics, 27(7). https://doi.org/10.1002/acm2.70673
Vancouver
Asahara T, Nishigami R, Kimoto N, Mitani M, Morimitsu Y, Akagi N, et al. Potential of quantitative X‐ray imaging from photon‐counting CT: A novel analysis based on effective atomic number and physical density. Journal of Applied Clinical Medical Physics. 2026;27(7). doi:10.1002/acm2.70673
BibTeX
@article{Asahara2026,
author = {Asahara, Takashi and Nishigami, Rina and Kimoto, Natsumi and Mitani, Mana and Morimitsu, Yusuke and Akagi, Noriaki and Higaki, Fumiyo and Iguchi, Toshihiro and Hiraki, Takao and Hayashi, Hiroaki},
title = {Potential of quantitative X‐ray imaging from photon‐counting CT: A novel analysis based on effective atomic number and physical density},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {7},
doi = {10.1002/acm2.70673}
}
EUVIMED