Development and validation of a 3D printed phantom for image quality assessment in fluoroscopy

Túlio Guilherme Soares Marques iD, Diana Rodrigues de Pina iD

Journal of Applied Clinical Medical Physics · 2026-06 · 27 (6)

DOI: 10.1002/acm2.70650 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 04.06.2026. Verfasst von Túlio Guilherme Soares Marques und 1 weiteren Autor:innen. Beteiligte Einrichtung: Institute of Bioscience, São Paulo State University (UNESP), Botucatu, São Paulo, Brazil. Hauptthemen laut MEDLINE: Phantoms, Imaging, Printing, Three-Dimensional, Computer-Aided Design, Image Processing, Computer-Assisted. Die Arbeit nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: São Paulo Research Foundation, Brazilian National Council for Scientific and Technological Development, Fundação de Amparo à Pesquisa do Estado de São Paulo.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Low‐contrast detectability is a key image‐quality parameter in fluoroscopy, directly influencing diagnostic accuracy and patient safety. Commercial phantoms such as the CDRAD provide reliable assessments but are costly and lack flexibility for routine use.

Purpose

To develop and validate a low‐cost 3D‐printed phantom for fluoroscopy quality control, designed with low‐contrast performance similar to the commercial reference standard CDRAD, and to offer a reproducible framework for clinical implementation. This represents an original and accessible strategy for systematic fluoroscopy image‐quality assessment.

Methods

The proposed 3D‐printed low‐contrast phantom (hereafter referred to as the LC phantom) was modeled in a Computer‐Aided Design (CAD) software and printed in polylactic acid (PLA) with 100% infill. It features a modular circular design adaptable to different fields of view. Quantitative validation included correlation between theoretical subject contrast (SC) and image‐based contrast‐to‐noise ratio (CNR), with comparisons to the CDRAD under clinically representative beam qualities (65–100 kVp), using additional copper filtration to produce hardened X‐ray spectra typical of fluoroscopic practice. Subjective comparison of both phantoms was performed through multicenter testing on four fluoroscopy systems (three mobile and one fixed).

Results

CNR increased consistently with hole depth for both phantoms across all beam qualities, with linear regression models yielding coefficients of determination (R 2 ) ranging from 0.88 to 0.99. Visual evaluations yielded mean low‐contrast thresholds of 3.11 ± 0.26%, 2.66 ± 0.31%, and 2.70 ± 0.27% for 65, 80, and 100 kVp, respectively, with a maximum deviation of 0.10 SC units relative to the reference phantom, indicating reproducible and consistent performance across the evaluated beam qualities.

Conclusions

The proposed LC phantom is a practical and affordable tool for routine fluoroscopy image quality assessment. Its methodology can be adapted to different beam qualities and institutional protocols, supporting harmonized and accessible image‐quality assessment in clinical practice.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingPrinting, Three-DimensionalComputer-Aided DesignImage Processing, Computer-AssistedQuality Assurance, Health CareFluoroscopyHumansQuality ControlSoftware

Stichwörter der Autor:innen

3D‐printed phantomfluoroscopyimage qualityinterventional radiologyquality control

Autor:innen und Institutionen

  • Institute of Bioscience, São Paulo State University (UNESP), Botucatu, São Paulo, Brazil.
  • Department of Tropical Diseases and Imaging Diagnosis, Faculty of Medicine of Botucatu, São Paulo State University (UNESP), Botucatu, São Paulo, Brazil.

ORCID-Profile: Túlio Guilherme Soares Marques, Diana Rodrigues de Pina

Förderung

  • São Paulo Research Foundation
  • Brazilian National Council for Scientific and Technological Development
  • Fundação de Amparo à Pesquisa do Estado de São Paulo

Zitieren

APA

Marques, T. G. S., & Pina, D. R. (2026). Development and validation of a 3D printed phantom for image quality assessment in fluoroscopy. Journal of Applied Clinical Medical Physics, 27(6). https://doi.org/10.1002/acm2.70650

Vancouver

Marques TGS, Pina DR. Development and validation of a 3D printed phantom for image quality assessment in fluoroscopy. Journal of Applied Clinical Medical Physics. 2026;27(6). doi:10.1002/acm2.70650

BibTeX
@article{Marques2026,
  author = {Marques, Túlio Guilherme Soares and Pina, Diana Rodrigues de},
  title = {Development and validation of a 3D printed phantom for image quality assessment in fluoroscopy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {6},
  doi = {10.1002/acm2.70650}
}
Kostenlos starten →