EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Adapt‐on‐demand: Enabling flexible and scalable adaptive radiotherapy through workflow automation

Sean Domal iD, Min Geon Choi, Justin Visak, Ruiqi Li iD, Taoran Li, Boon-Keng Kevin Teo, Xiaodong Zhao iD, Matthew Schmidt iD, Eric Laugeman, David Parsons iD, Yang Kyun Park, Shahed Badiyan und 2 weitere

Journal of Applied Clinical Medical Physics · 2026

DOI: 10.1002/acm2.70612 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Sean Domal und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, UT Southwestern Medical Center, Dallas, Texas, USA. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Workflow, Radiotherapy, Image-Guided, Neoplasms. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Varian Medical Systems, Inc., Department of Radiation Oncology at UT Southwestern Medical Center, Varian Medical Systems.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

CT‐guided adaptive radiotherapy (CTgART) offers significant clinical benefits, but its utilization is often limited by resource demands and rigid workflows. This study presents a flexible and scalable “Adapt‐on‐Demand” (AOD) framework that integrates the adaptive capabilities of the Varian Ethos platform with the high‐throughput IGRT delivery of the Varian Halcyon, supported by automated scripting.

Methods

We developed an end‐to‐end dual‐platform workflow that dynamically transitions between ART and IGRT delivery based on patient needs. Key components include Eclipse Scripting API (ESAPI) automation to streamline adapted plan ingestion from the Ethos Treatment Planning and Management System (TPMS) export directory, compatibility conversion for Halcyon delivery, and ARIA chart setup. The framework was evaluated through time‐efficiency analysis and longitudinal tracking of ART utilization and machine throughput. Multi‐institutional deployment was also conducted to validate generalizability.

Results

Automation reduced plan setup time from 45 to 8 min ( p < 0.001, unpaired t ‐test), and direct comparison of MLC leaf positions confirmed exact agreement between Ethos‐native and converted plans across all control points. ART utilization on the Ethos platform increased from 17% to 77% following IGRT offloading to Halcyon. The workflow was successfully implemented at two external sites without code modification, demonstrating portability.

Conclusion

This AOD framework enables flexible ART delivery without compromising workflow efficiency. By automating inter‐platform transitions and alleviating adaptive system burden, it improves clinical throughput and broadens patient access to ART. This work also underscores the importance of system interoperability and provides a scalable model adaptable across diverse practice settings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedWorkflowRadiotherapy, Image-GuidedNeoplasmsSoftwareHumansRadiotherapy DosageAutomationRadiotherapy, Intensity-ModulatedTomography, X-Ray ComputedAlgorithmsOrgans at RiskImage Processing, Computer-Assisted

Stichwörter der Autor:innen

CT‐guided adaptive radiotherapyESAPIEthosHalcyonadaptive radiotherapyadapt‐on‐demandworkflow automation

Autor:innen und Institutionen

  • Department of Radiation Oncology, UT Southwestern Medical Center, Dallas, Texas, USA.
  • Department of Radiation Oncology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
  • Department of Radiation Oncology, Washington University in St. Louis, St. Louis, Missouri, USA.

ORCID-Profile: Sean Domal, Ruiqi Li, Xiaodong Zhao, Matthew Schmidt, David Parsons

Förderung

  • Varian Medical Systems, Inc.
  • Department of Radiation Oncology at UT Southwestern Medical Center
  • Varian Medical Systems

Zitieren

APA

Domal, S., Choi, M. G., Visak, J., Li, R., Li, T., Teo, B. K., Zhao, X., Schmidt, M., Laugeman, E., Parsons, D., Park, Y. K., Badiyan, S., Sher, D., & Lin, M. H. (2026). Adapt‐on‐demand: Enabling flexible and scalable adaptive radiotherapy through workflow automation. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70612

Vancouver

Domal S, Choi MG, Visak J, Li R, Li T, Teo BK, et al. Adapt‐on‐demand: Enabling flexible and scalable adaptive radiotherapy through workflow automation. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70612

BibTeX
@article{Domal2026,
  author = {Domal, Sean and Choi, Min Geon and Visak, Justin and Li, Ruiqi and Li, Taoran and Teo, Boon-Keng Kevin and Zhao, Xiaodong and Schmidt, Matthew and Laugeman, Eric and Parsons, David and Park, Yang Kyun and Badiyan, Shahed and Sher, David and Lin, Mu-Han},
  title = {Adapt‐on‐demand: Enabling flexible and scalable adaptive radiotherapy through workflow automation},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70612}
}