Adapt‐on‐demand: Enabling flexible and scalable adaptive radiotherapy through workflow automation
Journal of Applied Clinical Medical Physics · 2026
DOI: 10.1002/acm2.70612 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Sean Domal und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, UT Southwestern Medical Center, Dallas, Texas, USA. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Workflow, Radiotherapy, Image-Guided, Neoplasms. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Varian Medical Systems, Inc., Department of Radiation Oncology at UT Southwestern Medical Center, Varian Medical Systems.
Abstract (Originalsprache)
Purpose
CT‐guided adaptive radiotherapy (CTgART) offers significant clinical benefits, but its utilization is often limited by resource demands and rigid workflows. This study presents a flexible and scalable “Adapt‐on‐Demand” (AOD) framework that integrates the adaptive capabilities of the Varian Ethos platform with the high‐throughput IGRT delivery of the Varian Halcyon, supported by automated scripting.
Methods
We developed an end‐to‐end dual‐platform workflow that dynamically transitions between ART and IGRT delivery based on patient needs. Key components include Eclipse Scripting API (ESAPI) automation to streamline adapted plan ingestion from the Ethos Treatment Planning and Management System (TPMS) export directory, compatibility conversion for Halcyon delivery, and ARIA chart setup. The framework was evaluated through time‐efficiency analysis and longitudinal tracking of ART utilization and machine throughput. Multi‐institutional deployment was also conducted to validate generalizability.
Results
Automation reduced plan setup time from 45 to 8 min ( p < 0.001, unpaired t ‐test), and direct comparison of MLC leaf positions confirmed exact agreement between Ethos‐native and converted plans across all control points. ART utilization on the Ethos platform increased from 17% to 77% following IGRT offloading to Halcyon. The workflow was successfully implemented at two external sites without code modification, demonstrating portability.
Conclusion
This AOD framework enables flexible ART delivery without compromising workflow efficiency. By automating inter‐platform transitions and alleviating adaptive system burden, it improves clinical throughput and broadens patient access to ART. This work also underscores the importance of system interoperability and provides a scalable model adaptable across diverse practice settings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, UT Southwestern Medical Center, Dallas, Texas, USA.
- Department of Radiation Oncology, University of Pennsylvania, Philadelphia, Pennsylvania, USA.
- Department of Radiation Oncology, Washington University in St. Louis, St. Louis, Missouri, USA.
ORCID-Profile: Sean Domal, Ruiqi Li, Xiaodong Zhao, Matthew Schmidt, David Parsons
Förderung
- Varian Medical Systems, Inc.
- Department of Radiation Oncology at UT Southwestern Medical Center
- Varian Medical Systems
Zitieren
APA
Domal, S., Choi, M. G., Visak, J., Li, R., Li, T., Teo, B. K., Zhao, X., Schmidt, M., Laugeman, E., Parsons, D., Park, Y. K., Badiyan, S., Sher, D., & Lin, M. H. (2026). Adapt‐on‐demand: Enabling flexible and scalable adaptive radiotherapy through workflow automation. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70612
Vancouver
Domal S, Choi MG, Visak J, Li R, Li T, Teo BK, et al. Adapt‐on‐demand: Enabling flexible and scalable adaptive radiotherapy through workflow automation. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70612
BibTeX
@article{Domal2026,
author = {Domal, Sean and Choi, Min Geon and Visak, Justin and Li, Ruiqi and Li, Taoran and Teo, Boon-Keng Kevin and Zhao, Xiaodong and Schmidt, Matthew and Laugeman, Eric and Parsons, David and Park, Yang Kyun and Badiyan, Shahed and Sher, David and Lin, Mu-Han},
title = {Adapt‐on‐demand: Enabling flexible and scalable adaptive radiotherapy through workflow automation},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70612}
}
EUVIMED