Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Oznur Senkesen und 3 weiteren Autor:innen. Beteiligte Einrichtung: Acıbadem Mehmet Ali Aydınlar University, Istanbul, Türkiye. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated, Organs at Risk, Unilateral Breast Neoplasms. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
This study aimed to dosimetrically evaluate the plan quality of a knowledge‐based planning (KBP) model developed from intensity‐modulated radiation therapy (IMRT) plans for left‐sided breast cancer treated with deep inspiration breath hold (DIBH), and to assess its clinical applicability. Materials and methods Eighty patients previously treated with DIBH were included. A RapidPlan‐based KBP model was developed using clinically approved plans (ClinP) within a Halcyon‐equivalent workflow on the Ethos platform. Dosimetric performance of model‐only plans (KBP) and plans refined with limited manual optimization (KBP+MP) was compared against ClinP for planning target volume (PTV) coverage and organ‐at‐risk (OAR) sparing.
Results
The KBP model demonstrated satisfactory training performance, with acceptable coefficients of determination (R 2 ) and mean squared error (MSE) values for most OARs. Both KBP and KBP+MP achieved PTV coverage comparable to ClinP, particularly for D95 and D98. In the left lung, V20 and D2% were reduced, and both the mean dose and D2% to the left anterior descending (LAD) artery were significantly lower in KBP plans, although lung V5 showed a slight increase.
Discussion
Our findings support the clinical applicability of KBP. While KBP effectively reduces high‐dose regions, it may increase low‐dose exposure particularly in distant OARs such as the contralateral lung and breast. The KBP+MP approach provided more balanced OAR sparing while simplifying the overall planning workflow compared with manual methods.
Conclusion
The KBP model demonstrated clinical feasibility for left‐sided breast IMRT under DIBH conditions. Model‐generated plans achieved target coverage comparable to clinically approved plans while providing acceptable OAR sparing. Although KBP alone tended to increase low‐dose exposure in contralateral structures, limited manual refinement improved overall dose balance. These findings support the integration of structured KBP workflows into standardized DIBH breast IMRT planning.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Acıbadem Mehmet Ali Aydınlar University, Istanbul, Türkiye.
- Department of Radiation Oncology, Acıbadem Ataşehir Hospital, İstanbul, Türkiye.
- Department of Radiation Oncology, Acıbadem Altunizade Hospital, İstanbul, Türkiye.
ORCID-Profile: Oznur Senkesen, Emine Burcin Ispir, Irem Aydin, Evrim Tezcanli
Zitieren
APA
Senkesen, O., Ispir, E. B., Aydin, I., & Tezcanli, E. (2026). Implementation and Dosimetric Evaluation of a Knowledge‐Based IMRT Model for Left‐Sided Breast Radiotherapy with DIBH. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70599
Vancouver
Senkesen O, Ispir EB, Aydin I, Tezcanli E. Implementation and Dosimetric Evaluation of a Knowledge‐Based IMRT Model for Left‐Sided Breast Radiotherapy with DIBH. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70599
BibTeX
@article{Senkesen2026,
author = {Senkesen, Oznur and Ispir, Emine Burcin and Aydin, Irem and Tezcanli, Evrim},
title = {Implementation and Dosimetric Evaluation of a Knowledge‐Based IMRT Model for Left‐Sided Breast Radiotherapy with DIBH},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
doi = {10.1002/acm2.70599}
}
EUVIMED