Image quality comparison of synthesized two‐dimensional images and digital mammography in breast tomosynthesis: A quantitative analysis

Kazumi Sogabe, Sho Maruyama, Nao Koyama, Hiroki Saito iD

Journal of Applied Clinical Medical Physics · 2026-04 · 27 (4)

DOI: 10.1002/acm2.70574 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 09.04.2026. Verfasst von Kazumi Sogabe und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Sciences, School of Health Sciences, Ibaraki Prefectural University of Health Sciences, Ami, Ibaraki, Japan. Hauptthemen laut MEDLINE: Mammography, Phantoms, Imaging, Breast Neoplasms, Radiographic Image Enhancement. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Breast cancer remains a leading cause of cancer‐related morbidity and mortality worldwide. Early detection of breast cancer using high‐quality mammography is essential for improving prognosis and treatment outcomes. Digital breast tomosynthesis with synthesized two‐dimensional mammography (SM) reduces dose and tissue overlap; however, its comprehensive physical evaluations remain limited.

Purpose

This study quantitatively compared the image quality of conventional digital mammography (DM) and SM acquired using the AMULET Innovality System.

Methods

Polymethyl methacrylate, American College of Radiology, and contrast‐detail (CD) mammography phantoms were imaged 10 times in both DM and SM modes. Image quality metrics, including signal‐to‐noise ratio (SNR) maps, noise power spectrum (NPS), contrast‐to‐noise ratio (CNR), modulation transfer function (MTF), and CD curves, were evaluated.

Results

DM demonstrated higher SNR uniformity and suppressed low‐ and high‐frequency noise, and the CNR values were 1.3–2 times greater than those of SM across all object sizes. MTF analysis showed superior resolution in DM (MTF(f50), 4.0 cycles/mm) compared with that of the SM (2.5 cycles/mm). CD curve analysis confirmed the better detectability of fine structures in DM, with values of 134.9 for DM and 56.7 for SM.

Conclusions

Quantitative phantom‐based evaluation demonstrated that DM showed superior performance to SM in several physical image quality metrics, including SNR uniformity, noise characteristics, spatial resolution, and contrast–detail detectability. These findings indicate systematic differences in physical image quality between DM and SM; however, further observer and clinical studies are required to clarify how these differences translate to diagnostic performance in clinical practice.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

MammographyPhantoms, ImagingBreast NeoplasmsRadiographic Image EnhancementImage Processing, Computer-AssistedRadiographic Image Interpretation, Computer-AssistedHumansFemaleSignal-To-Noise RatioAlgorithms

Stichwörter der Autor:innen

breast cancerconventional digital mammographysynthesized two‐dimensional mammography

Autor:innen und Institutionen

  • Department of Radiological Sciences, School of Health Sciences, Ibaraki Prefectural University of Health Sciences, Ami, Ibaraki, Japan.
  • Department of Radiological Technology, Gunma Prefectural College of Health Sciences, Maebashi, Gunma, Japan.
  • Department of Radiology, Teikyo University Hospital, Itabashi, Tokyo, Japan.
  • Department of Medical Radiology, Faculty of Medical Technology, Teikyo University, Itabashi, Tokyo, Japan.

ORCID-Profile: Hiroki Saito

Zitieren

APA

Sogabe, K., Maruyama, S., Koyama, N., & Saito, H. (2026). Image quality comparison of synthesized two‐dimensional images and digital mammography in breast tomosynthesis: A quantitative analysis. Journal of Applied Clinical Medical Physics, 27(4). https://doi.org/10.1002/acm2.70574

Vancouver

Sogabe K, Maruyama S, Koyama N, Saito H. Image quality comparison of synthesized two‐dimensional images and digital mammography in breast tomosynthesis: A quantitative analysis. Journal of Applied Clinical Medical Physics. 2026;27(4). doi:10.1002/acm2.70574

BibTeX
@article{Sogabe2026,
  author = {Sogabe, Kazumi and Maruyama, Sho and Koyama, Nao and Saito, Hiroki},
  title = {Image quality comparison of synthesized two‐dimensional images and digital mammography in breast tomosynthesis: A quantitative analysis},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  volume = {27},
  number = {4},
  doi = {10.1002/acm2.70574}
}
Kostenlos starten →