Quantitative evaluation of the effect of anatomical changes on the results of in vivo dosimetry during radiation therapy in the thoracic region
Journal of Applied Clinical Medical Physics · 2026-04 · 27 (4)
DOI: 10.1002/acm2.70542 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 25.03.2026. Verfasst von Kaname Tanaka und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Therapy, Toyota Memorial Hospital, Toyota, Aichi, Japan. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated, Organs at Risk, Lung Neoplasms. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Anatomical changes during thoracic radiation therapy can alter dose delivery and compromise accuracy. Electronic portal imaging device (EPID)‐based in vivo dosimetry (IVD) enables noninvasive monitoring of these changes during treatment.
Purpose
This study quantitatively evaluated the impact of EPID‐based IVD in volumetric modulated arc therapy (VMAT) and explored the potential of IVD for guiding timely replanning.
Methods
Eleven patients undergoing thoracic VMAT were retrospectively analyzed: six required replanning as a result of anatomical changes (replanning group), and five did not (normal group). For each fraction, cumulative EPID transit images were acquired, and the global gamma passing rate (GPR) was calculated using six criteria. The difference between treatment GPR and patient‐specific QA GPR (ΔGPR) was used to reduce plan‐specific bias. For four patients in the replanning group (excluding two patients who were excluded due to changes in treatment machines or immobilization), the original plans were recalculated on post‐replanning CT images to estimate dose changes in the planning target volume (PTV) and organs at risks (OARs).
Results
Across all criteria, the normal group demonstrated higher GPR and ΔGPR values. The 3%/3 mm criterion best discriminated between groups, with a per‐fraction GPR of 98.2% ± 1.1% (normal) versus 90.5% ± 7.1% (replanning). GPR improved after replanning but remained below the levels observed in the normal group. Dose simulations indicated that continuing treatment without replanning increased the PTV dose (D50% up to + 3.5%) and OAR doses, with increases exceeding 5 Gy in some cases.
Conclusions
EPID‐based IVD detected reduced GPR values in replanning cases, identifying the 3%/3 mm criterion as optimal. A per‐fraction GPR of < 97% under this criterion may indicate notable anatomical changes. Continuing treatment without replanning tended to increase doses delivered to the target and normal tissues. These findings support the clinical utility of IVD and provide quantitative criteria for replanning decisions beyond CT or contour‐based assessments.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Therapy, Toyota Memorial Hospital, Toyota, Aichi, Japan.
- Department of Integrated Health Sciences, Nagoya University Graduate School of Medicine, Nagoya, Aichi, Japan.
Zitieren
APA
Tanaka, K., & Komori, M. (2026). Quantitative evaluation of the effect of anatomical changes on the results of in vivo dosimetry during radiation therapy in the thoracic region. Journal of Applied Clinical Medical Physics, 27(4). https://doi.org/10.1002/acm2.70542
Vancouver
Tanaka K, Komori M. Quantitative evaluation of the effect of anatomical changes on the results of in vivo dosimetry during radiation therapy in the thoracic region. Journal of Applied Clinical Medical Physics. 2026;27(4). doi:10.1002/acm2.70542
BibTeX
@article{Tanaka2026,
author = {Tanaka, Kaname and Komori, Masataka},
title = {Quantitative evaluation of the effect of anatomical changes on the results of in vivo dosimetry during radiation therapy in the thoracic region},
journal = {Journal of Applied Clinical Medical Physics},
year = {2026},
volume = {27},
number = {4},
doi = {10.1002/acm2.70542}
}