Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2026. Verfasst von Abdulaziz Alhussan und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Medicine, Oregon Health and Science University, Portland, Oregon, USA. Hauptthemen laut MEDLINE: Electrons, Phantoms, Imaging, Particle Accelerators, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Kuwait Foundation for the Advancement of Sciences.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

This study evaluates the dosimetric impact of integrating thin metallic scatter foils with Cerrobend contact skin collimators to improve dose uniformity, conformality, and distal tissue sparing in small superficial electron fields. Materials and methods Electron beams of 8, 12, and 15 MeV from an Elekta Versa HD LINAC were delivered through a Cerrobend skin collimator with a 2.0 cm aperture at 100.0 cm SSD. Thin aluminum (Al) and lead (Pb) foils (< 2.50 mm) were placed on the collimator. Gafchromic EBT3 film in a solid‐water phantom was used to measure depth‐dose distributions and isodose profiles following TG‐235–consistent calibration.

Results

Scatter foils produced thickness‐ and Z‐dependent modulation of beam characteristics. At 8 MeV, the thickest Pb foil (1.07 mm) reduced the practical range ( R p ) by ∼45% and shifted D m a x proximally by ∼0.5 cm, yielding substantial distal tissue sparing. Al foils caused smaller R p reductions (15%–25%) but improved lateral dose uniformity, producing smoother and more symmetric isodose contours. The 90% isodose diameter decreased with foil thickness for both materials, with Pb showing the largest contraction (∼20%–25%, energy‐dependent), enhancing field conformality. Penumbra width increased slightly for thin foils but stabilized or narrowed for larger thicknesses. These effects diminished at 15 MeV, indicating reduced sensitivity of high‐energy electrons to thin‐foil perturbation.

Conclusions

Thin metallic foils placed on Cerrobend skin collimators enable a controllable balance between dose uniformity (improved with Al) and conformality with distal sparing (enhanced with Pb). This simple, LINAC‐independent configuration offers a cost‐effective method for modulating small‐field electron beam characteristics and may serve as a practical adjunct for treating superficial lesions.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectronsPhantoms, ImagingParticle AcceleratorsRadiotherapy Planning, Computer-AssistedNeoplasmsHumansRadiotherapy DosageScattering, RadiationAluminumRadiometryLeadRadiotherapy, Intensity-ModulatedOrgans at Risk

Stichwörter der Autor:innen

cerrobendelectron beam therapyisodose distributionpenumbrascatter foilskin collimatorsmall field dosimetrysuperficial tumors

Substanzen

AluminumLead

Autor:innen und Institutionen

  • Department of Radiation Medicine, Oregon Health and Science University, Portland, Oregon, USA.

Förderung

  • Kuwait Foundation for the Advancement of Sciences

Zitieren

APA

Alhussan, A., & Crilly, R. (2026). Dosimetric characterization of scatter foil‐enhanced contact collimation for small superficial electron beam therapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70484

Vancouver

Alhussan A, Crilly R. Dosimetric characterization of scatter foil‐enhanced contact collimation for small superficial electron beam therapy. Journal of Applied Clinical Medical Physics. 2026. doi:10.1002/acm2.70484

BibTeX
@article{Alhussan2026,
  author = {Alhussan, Abdulaziz and Crilly, Richard},
  title = {Dosimetric characterization of scatter foil‐enhanced contact collimation for small superficial electron beam therapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2026},
  doi = {10.1002/acm2.70484}
}
Kostenlos starten →