Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Evren Ozan Goksel und 8 weiteren Autor:innen. Beteiligte Einrichtung: Vocational School of Health Services, Radiotherapy Program, Acibadem MAA University, Istanbul, Turkey. Hauptthemen laut MEDLINE: Radiosurgery, Lung Neoplasms, Radiotherapy Planning, Computer-Assisted, Carcinoma, Non-Small-Cell Lung. Die Arbeit nennt 56 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To report the initial clinical experience with the radixact synchrony lung with respiratory (LWR) technique in stereotactic body radiotherapy (SBRT) for lung tumors, focusing on feasibility, dosimetric performance, and early clinical outcomes using both standard and extended fractionation regimens.

Methods

Twenty patients with primary or metastatic lung tumors were treated using the Synchrony LWR technique. Patients were selected based on respiratory stability and favorable imaging characteristics. All patients underwent a Synchrony simulation, which included tumor tracking volume (TTV) definition and kV radiograph acquisition to assess their suitability for treatment with this technique. Treatment plans were generated using Precision TPS with VoloUltra. Plan verification, including gamma analysis and point dose measurements, was performed using the PTW Octavius system and dynamic motion QA. Synchrony model parameters, treatment times, and intrafraction interruptions were analyzed. Clinical follow‐up included imaging and toxicity assessment.

Results

Treatment was successfully delivered in all patients with median beam‐on and treatment durations of 9.1 and 14.5 min, respectively. The median number of intrafraction interruptions was 1.95, and none of the fractions were canceled due to prolonged delivery. The mean kV imaging dose per fraction was 0.90 cGy. Dosimetric QA under motion showed gamma passing rates >93% and < 2% deviation in point dose. At a median follow‐up of 9 months, the local control rate was 100%, with no grade ≥2 toxicities observed.

Conclusions

The Radixact Synchrony LWR technique enables efficient and non‐invasive respiratory motion management in lung SBRT. This early experience demonstrates its feasibility with clinically acceptable delivery accuracy and promising short‐term clinical outcomes.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryLung NeoplasmsRadiotherapy Planning, Computer-AssistedCarcinoma, Non-Small-Cell LungHumansRadiotherapy DosageMaleFemaleAgedRadiotherapy, Intensity-ModulatedMiddle AgedAged, 80 and overOrgans at RiskDose Fractionation, RadiationPrognosisFollow-Up Studies

Stichwörter der Autor:innen

lung SBRTmarkerless tumor trackingmotion managementradixact synchrony

Autor:innen und Institutionen

  • Vocational School of Health Services, Radiotherapy Program, Acibadem MAA University, Istanbul, Turkey.
  • Department of Radiation Oncology, Acibadem Altunizade Hospital, Istanbul, Turkey.
  • Department of Radiation Oncology, Istanbul Onkoloji Hospital, Istanbul, Turkey.
  • Department Of Medical Sciences, School Of Medicine, Radiation Oncology, Acibadem MAA University, Istanbul, Turkey.

ORCID-Profile: Evren Ozan Goksel

Zitieren

APA

Goksel, E. O., Ozen, Z., Gunduz, S., Ture, A., Kucucuk, H., Arifoglu, A., Tezcanli, E., Abacioglu, U., & Sengoz, M. (2025). Markerless tumor tracking for lung SBRT with the Radixact Synchrony system: Initial experience and short‐term clinical outcomes. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70362

Vancouver

Goksel EO, Ozen Z, Gunduz S, Ture A, Kucucuk H, Arifoglu A, et al. Markerless tumor tracking for lung SBRT with the Radixact Synchrony system: Initial experience and short‐term clinical outcomes. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70362

BibTeX
@article{Goksel2025,
  author = {Goksel, Evren Ozan and Ozen, Zeynep and Gunduz, Seckin and Ture, Artunc and Kucucuk, Halil and Arifoglu, Alptekin and Tezcanli, Evrim and Abacioglu, Ufuk and Sengoz, Meric},
  title = {Markerless tumor tracking for lung SBRT with the Radixact Synchrony system: Initial experience and short‐term clinical outcomes},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70362}
}
Kostenlos starten →