EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Automatic radiotherapy planning for deliverable plans using deep learning dose prediction and dose rings optimization in cervical cancer

Weiqian Huang, Ting Liu, Yichao Shen, Ziqing Xiang, Dong Wang, Wen Fu, Li Shao, Xianwen Yu, Weihua Ni, Yongqiang Zhou, Huan Liu, Ce Han und 2 weitere

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70353 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Weiqian Huang und 13 weiteren Autor:innen. Beteiligte Einrichtung: The 1st School of Medicine (School of information and Engineering), Wenzhou Medical University, Zhejiang, China. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Deep Learning, Radiotherapy, Intensity-Modulated, Organs at Risk. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Zhejiang Natural Science Foundation, Key project of Zhejiang Provincial Health Science and Technology Program, Major project of Wenzhou Science and Technology Bureau.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Automatic radiotherapy (RT) planning based on deep learning (DL) has been extensively researched. However, it is challenging to import the predicted dose distribution into mainstream treatment planning systems (TPSs) and generate clinically deliverable plans.

Purpose

To investigate the feasibility and accuracy of an automatic volumetric modulated arc therapy (VMAT) and intensity‐modulated radiation therapy (IMRT) planning method for generation of universally deliverable plans based on DL dose prediction and dose rings optimization.

Methods

First, dose distributions were predicted using a three‐dimensional (3D) Fusion Residual Unet (F‐ResUnet) DL network with data from two hospitals, which included 230 and 210 gynecological cancer (GC) patients underwent VMAT and IMRT, respectively. Then, the predicted dose distributions were discretized into dose rings to optimize the plans automatically in two mainstream TPSs based on the dose rings. Finally, the deliverability of generated plans was verified with patient‐specific quality assurance (PSQA).

Results

The predicted dose distributions were clinically acceptable with a target coverage over 95%. Compared with the clinical plans, the automatic plans optimized with dose rings achieved a similar dose coverage on planning target volumes (PTV) with an average target coverage over 96.5%. For organs at risk (OARs) sparing, automatic VMAT plans markedly decreased the V 30Gy of left femoral head ( p = 0.05), right femoral head ( p = 0.004), and small intestine ( p = 0.04). The V 45Gy of bladder and rectum in the automatic IMRT plans were reduced by approximately 7% and 9%, respectively. Deliverability verification with PSQA achieved a mean gamma passing rate of 99.1%, 97.1% and 98.3%, 95.0% under the criteria of 3%/3 mm and 3%/2 mm for VMAT and IMRT plans, respectively.

Conclusions

The proposed automatic planning method combining DL dose prediction and dose rings optimization was feasible to generate universally deliverable VMAT and IMRT plans for gynecological cancer (GC) patients.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedDeep LearningRadiotherapy, Intensity-ModulatedOrgans at RiskUterine Cervical NeoplasmsHumansFemaleRadiotherapy DosageQuality Assurance, Health Care

Stichwörter der Autor:innen

automatic planningdeep learningdose rings optimizationintensity‐modulated radiation therapyvolumetric modulated arc therapy

Autor:innen und Institutionen

  • The 1st School of Medicine (School of information and Engineering), Wenzhou Medical University, Zhejiang, China.
  • Department of Radiation Oncology, Taizhou Hospital, Taizhou, China.
  • Department of Radiotherapy and Medical Oncology, 1st Affiliated Hospital of Wenzhou Medical University, Wenzhou, China.
  • Department of Medical Engineering and Equipment, 2nd Affiliated Hospital of Wenzhou Medical University, Wenzhou, China.
  • Cixi Biomedical Research Institute, Wenzhou Medical University, Zhejiang, China.
  • School of Basic Medical Science, Wenzhou Medical University, Wenzhou, China.

ORCID-Profile: Xiance Jin

Förderung

  • Zhejiang Natural Science Foundation
  • Key project of Zhejiang Provincial Health Science and Technology Program
  • Major project of Wenzhou Science and Technology Bureau
  • Project Supported by Scientific Research Fund of Zhejiang Provincial Education Department
  • Project of Wenzhou Science and Technology Bureau

Zitieren

APA

Huang, W., Liu, T., Shen, Y., Xiang, Z., Wang, D., Fu, W., Shao, L., Yu, X., Ni, W., Zhou, Y., Liu, H., Han, C., Jin, X., & Zhang, J. (2025). Automatic radiotherapy planning for deliverable plans using deep learning dose prediction and dose rings optimization in cervical cancer. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70353

Vancouver

Huang W, Liu T, Shen Y, Xiang Z, Wang D, Fu W, et al. Automatic radiotherapy planning for deliverable plans using deep learning dose prediction and dose rings optimization in cervical cancer. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70353

BibTeX
@article{Huang2025,
  author = {Huang, Weiqian and Liu, Ting and Shen, Yichao and Xiang, Ziqing and Wang, Dong and Fu, Wen and Shao, Li and Yu, Xianwen and Ni, Weihua and Zhou, Yongqiang and Liu, Huan and Han, Ce and Jin, Xiance and Zhang, Ji},
  title = {Automatic radiotherapy planning for deliverable plans using deep learning dose prediction and dose rings optimization in cervical cancer},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70353}
}
Kostenlos starten →