EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

A hybrid IGRT workflow using SGRT and CBCT for prostate SBRT: Feasibility, efficiency, and safety

Juergen Meyer iD, Angelia Tran, Ting Martin Ma iD, Bing-Hao Chiang iD, Tamara Egan, Jonathan J Chen, Yinghua Tao, Ning Cao iD, Katherine H Kim, Jay J Liao iD, Sharareh Koufigar iD, Winston Vuong iD und 1 weitere

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70339 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Juergen Meyer und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Fred Hutchinson Cancer Center, University of Washington, Seattle, Washington, USA. Hauptthemen laut MEDLINE: Prostatic Neoplasms, Cone-Beam Computed Tomography, Radiosurgery, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 53 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and purpose Safe delivery of prostate stereotactic body radiotherapy (SBRT) relies on precise target localization. Without access to real‐time intrafraction motion management, careful optimization of IGRT protocols is necessary to safeguard treatment accuracy and patient outcomes.

Methods

An IGRT workflow is proposed that incorporates surface‐monitoring (SGRT) to complement cone‐beam CT (CBCT) imaging. The study evaluates 23 consecutive SBRT prostate patients who were treated on a prospective registry study. Each patient received pre‐ and mid‐treatment and a subset received post‐treatment CBCTs. The frequency and magnitude of SGRT triggered beam interruptions as well as treatment times were recorded.

Results

The median number of CBCTs acquired per fraction was four and the median treatment time was 23 min (IQR 19–27). SGRT detected intra‐fraction surface‐based motion beyond a combined 4 mm vector isocenter tolerance in 62% of all fractions treated, with a maximum motion of 15 mm. On average < 2 beam interruptions were triggered by SGRT per treatment fraction. There was no statistically significant correlation between overall treatment time and SGRT‐triggered beam interruptions ( r = 0.048, p = 0.645). There was a weak but statistically relevant correlation of overall treatment time with the maximum detected motion ( r = 0.23, p = 0.026). SGRT detected five fractions where the patients had persistently moved outside the SGRT tolerance, and for three of these (60%), a CBCT verified that the target was out of tolerance.

Conclusion

SGRT is a valuable tool that complements CBCT‐based IGRT. An SGRT motion vector tolerance of 4 mm provides a pragmatic compromise between detecting patient motion and treatment efficiency. Overall, persistent patient motion during treatment was infrequent in this cohort, however, SGRT was able to detect several cases where the internal target was outside of the tolerance highlighting that patient monitoring with SGRT can contribute to improved quality and safety for prostate SBRT.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Prostatic NeoplasmsCone-Beam Computed TomographyRadiosurgeryRadiotherapy Planning, Computer-AssistedWorkflowRadiotherapy, Intensity-ModulatedRadiotherapy, Image-GuidedHumansMaleRadiotherapy DosageFeasibility StudiesProspective StudiesOrgans at RiskAgedImage Processing, Computer-Assisted

Stichwörter der Autor:innen

SGRT tolerancesimage guided radiation therapy (IGRT) workflow optimizationpatient monitoringpatient safetyprostate stereotactic body radiation therapy (SBRT)surface guided radiotherapy (SGRT)

Autor:innen und Institutionen

  • Department of Radiation Oncology, Fred Hutchinson Cancer Center, University of Washington, Seattle, Washington, USA.

ORCID-Profile: Juergen Meyer, Ting Martin Ma, Bing-Hao Chiang, Ning Cao, Jay J Liao, Sharareh Koufigar, Winston Vuong, Emily S Weg

Zitieren

APA

Meyer, J., Tran, A., Ma, T. M., Chiang, B. H., Egan, T., Chen, J. J., Tao, Y., Cao, N., Kim, K. H., Liao, J. J., Koufigar, S., Vuong, W., & Weg, E. S. (2025). A hybrid IGRT workflow using SGRT and CBCT for prostate SBRT: Feasibility, efficiency, and safety. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70339

Vancouver

Meyer J, Tran A, Ma TM, Chiang BH, Egan T, Chen JJ, et al. A hybrid IGRT workflow using SGRT and CBCT for prostate SBRT: Feasibility, efficiency, and safety. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70339

BibTeX
@article{Meyer2025,
  author = {Meyer, Juergen and Tran, Angelia and Ma, Ting Martin and Chiang, Bing-Hao and Egan, Tamara and Chen, Jonathan J and Tao, Yinghua and Cao, Ning and Kim, Katherine H and Liao, Jay J and Koufigar, Sharareh and Vuong, Winston and Weg, Emily S},
  title = {A hybrid IGRT workflow using SGRT and CBCT for prostate SBRT: Feasibility, efficiency, and safety},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70339}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.