Clinical evaluation of the aperture shape controller in volumetric modulated arc therapy: Effects on MLC and jaw motion complexity, plan quality, and deliverability
Journal of Applied Clinical Medical Physics · 2025
DOI: 10.1002/acm2.70215 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Tatsuya Kamima und 12 weiteren Autor:innen. Beteiligte Einrichtung: Radiation Oncology Department, Cancer Institute Hospital, Japanese Foundation for Cancer Research, Koto-ku, Tokyo, Japan. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated, Radiotherapy Planning, Computer-Assisted, Neoplasms, Jaw. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
This study comprehensively evaluated the impact of aperture shape controller (ASC) on volumetric modulated arc therapy (VMAT) clinical treatment planning.
Methods
A total of 248 VMAT plans for head and neck, prostate, post mastectomy breast, gastric MALT lymphoma, rectum, and Lung SBRT for treatments performed between 2018 and 2023 were retrospectively enrolled. These treatment plans were divided into two groups: before (ASC‐OFF group) and after (ASC‐ON group) the clinical implementation of ASC. The performance of ASC in terms of plan complexity was evaluated by between group comparisons of the multi‐leaf collimator (MLC) modulation complexity score (MCS v ) and jaw tracking complexity score (JTCS), plan quality using dosimetric parameters, and plan deliverability.
Results
The overall MCS v and JTCS of the ASC‐OFF and ASC‐ON groups were 0.318 and 0.369 ( p < 0.001) and 2.85 cm and 2.98 cm ( p = 0.10), respectively. In the ASC‐ON group, the complexity of MLC motions was significantly lower and the complexity of jaw motions higher. In addition, most dosimetric parameters were comparable between groups or better in the ASC‐ON group. There was no significant difference in the gamma passing rate of 3%/2 mm between the two groups, although at tighter criteria, the gamma passing rate was higher in the ASC‐ON group.
Conclusions
The use of ASC reduced the complexity of MLC motions without compromising the plan quality. Furthermore, this study found that ASC allows jaw tracking to be more effective.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Radiation Oncology Department, Cancer Institute Hospital, Japanese Foundation for Cancer Research, Koto-ku, Tokyo, Japan.
ORCID-Profile: Tatsuya Kamima, Masahiro Kaneko
Zitieren
APA
Kamima, T., Sato, Y., Murakami, Y., Takahashi, T., Kaneko, M., Watanabe, S., Miyauchi, H., Abo, N., Ito, Y., Toshiyasu, T., Taguchi, S., Harada, A., & Yoshioka, Y. (2025). Clinical evaluation of the aperture shape controller in volumetric modulated arc therapy: Effects on MLC and jaw motion complexity, plan quality, and deliverability. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70215
Vancouver
Kamima T, Sato Y, Murakami Y, Takahashi T, Kaneko M, Watanabe S, et al. Clinical evaluation of the aperture shape controller in volumetric modulated arc therapy: Effects on MLC and jaw motion complexity, plan quality, and deliverability. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70215
BibTeX
@article{Kamima2025,
author = {Kamima, Tatsuya and Sato, Yosuke and Murakami, Yuki and Takahashi, Taro and Kaneko, Masahiro and Watanabe, Shiori and Miyauchi, Hikaru and Abo, Natsumi and Ito, Yasushi and Toshiyasu, Takashi and Taguchi, Senzo and Harada, Arisa and Yoshioka, Yasuo},
title = {Clinical evaluation of the aperture shape controller in volumetric modulated arc therapy: Effects on MLC and jaw motion complexity, plan quality, and deliverability},
journal = {Journal of Applied Clinical Medical Physics},
year = {2025},
doi = {10.1002/acm2.70215}
}
EUVIMED