Implementation of a real‐time neutron monitor system for use in routine quality assurance of an accelerator‐based neutron system for clinical boron neutron capture therapy

Naonori Hu iD, Taiki Nakamura, Ryusuke Kataura, Keita Suga, Tetsuya Mukawa, Kazuhiko Akita, Ryo Kakino, Akinori Sasaki, Mai Nojiri iD, Nishiki Matsubayashi, Takushi Takata, Hiroki Tanaka und 2 weitere

Journal of Applied Clinical Medical Physics · 2025

DOI: 10.1002/acm2.70190 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2025. Verfasst von Naonori Hu und 13 weiteren Autor:innen. Beteiligte Einrichtung: Kansai BNCT Medical Center, Osaka Medical and Pharmaceutical University, Osaka, Japan. Hauptthemen laut MEDLINE: Boron Neutron Capture Therapy, Particle Accelerators, Quality Assurance, Health Care, Phantoms, Imaging. Die Arbeit nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Sumitomo Heavy Industries, Ltd..

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Currently, the metal foil activation method is routinely used to measure the neutron output of an accelerator‐based neutron source designed for clinical Boron neutron capture therapy (BNCT). Although this method is well established and has been primarily utilized since the nuclear reactor BNCT era, the process is labour‐intensive and not well‐suited for a busy hospital environment performing routine patient treatment. A replacement neutron detector system that is simple to use and can measure the neutron output in real‐time is necessary.

Purpose

Investigation and implementation of an Eu doped LiCaAlF 6 scintillator detector for use in routine quality assurance tests of an accelerator‐based neutron source designed for clinical BNCT.

Methods

The response of the scintillator detector was evaluated using the NeuCure BNCT system installed at the Kansai BNCT Medical Center. The measurement repeatability, neutron fluence linearity, and neutron flux dependency of the detector system were evaluated. The beam central axis and off‐axis thermal neutron distribution inside a water phantom were measured and compared with the Monte Carlo treatment planning system (TPS).

Results

The scintillator detector system showed high measurement repeatability with a coefficient of variation of less than 0.4%. The detector system showed linear response up to a proton charge of 3.6 C, and the response was stable between a proton current of 0.1 and 1 mA. Both the central axis and off‐axis thermal neutron flux inside a water phantom matched closely with both the metal foil activation method and the Monte Carlo simulation results. The time it took to perform a routine quality assurance test was drastically reduced from 1.5 h down to a few minutes.

Conclusion

Implementation of this detector system in the clinic would significantly reduce the time required for routine QA, acceptance, and commissioning, and be a stepping stone to assist expansion of accelerator‐based BNCT systems worldwide.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Boron Neutron Capture TherapyParticle AcceleratorsQuality Assurance, Health CarePhantoms, ImagingNeutronsRadiotherapy Planning, Computer-AssistedNeoplasmsHumansRadiotherapy DosageMonte Carlo MethodScintillation Counting

Stichwörter der Autor:innen

BNCTMonte Carlo simulationQAReal‐time neutron monitoraccelerator neutron source

Autor:innen und Institutionen

  • Kansai BNCT Medical Center, Osaka Medical and Pharmaceutical University, Osaka, Japan.
  • Institute for Integrated Radiation and Nuclear Science, Kyoto University, Osaka, Japan.
  • Sumitomo Heavy Industries, Ltd., Tokyo, Japan.
  • Graduate School of Engineering, Kyoto University, Kyoto, Japan.
  • Department of Radiation Oncology, Osaka Medical and Pharmaceutical University Hospital, Osaka, Japan.

ORCID-Profile: Naonori Hu, Mai Nojiri

Förderung

  • Sumitomo Heavy Industries, Ltd.

Zitieren

APA

Hu, N., Nakamura, T., Kataura, R., Suga, K., Mukawa, T., Akita, K., Kakino, R., Sasaki, A., Nojiri, M., Matsubayashi, N., Takata, T., Tanaka, H., Nihei, K., & Ono, K. (2025). Implementation of a real‐time neutron monitor system for use in routine quality assurance of an accelerator‐based neutron system for clinical boron neutron capture therapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.70190

Vancouver

Hu N, Nakamura T, Kataura R, Suga K, Mukawa T, Akita K, et al. Implementation of a real‐time neutron monitor system for use in routine quality assurance of an accelerator‐based neutron system for clinical boron neutron capture therapy. Journal of Applied Clinical Medical Physics. 2025. doi:10.1002/acm2.70190

BibTeX
@article{Hu2025,
  author = {Hu, Naonori and Nakamura, Taiki and Kataura, Ryusuke and Suga, Keita and Mukawa, Tetsuya and Akita, Kazuhiko and Kakino, Ryo and Sasaki, Akinori and Nojiri, Mai and Matsubayashi, Nishiki and Takata, Takushi and Tanaka, Hiroki and Nihei, Keiji and Ono, Koji},
  title = {Implementation of a real‐time neutron monitor system for use in routine quality assurance of an accelerator‐based neutron system for clinical boron neutron capture therapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2025},
  doi = {10.1002/acm2.70190}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche