Evaluation of the measurement accuracy and uncertainty of a solid‐state detector under diagnostic x‐ray beam conditions
Journal of Applied Clinical Medical Physics · 2024-09 · 25 (9)
DOI: 10.1002/acm2.14476 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 19.07.2024. Verfasst von Sho Maruyama und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Technology, Gunma Prefectural College of Health Sciences, Gunma, Japan. Hauptthemen laut MEDLINE: Radiation Dosage. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 27 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective
An accurate measurement of x‐ray beams is expected to reduce the uncertainties associated with estimating radiation risk to patients in clinical settings. To perform assessment tasks based on the readings of a solid‐state detector (SSD) using semiconductor technology, the characteristics of the detector should be elucidated. In this study, we evaluated the measurement accuracy of a new SSD under diagnostic x‐ray beam conditions in terms of air kerma, tube voltage, and half‐value layer (HVL). The performance of the SSD was then compared with those of reference instruments.
Methods
The tube voltage was varied within the range of 50–120 kV in steps of 10 kV and the thickness and materials of additional filters were concurrently changed (several combinations were tested). In addition, the dose rate and energy dependence of the SSD were also investigated. These effects were analyzed based on statistical significance tests. Furthermore, the expanded uncertainties in the series of measurements were meticulously calculated.
Results
The results showed average relative differences of −3.26 ± 1.33%, 0.44 ± 1.01%, and −2.60 ± 3.31% for air kerma, tube voltage, and HVL, respectively. Furthermore, air kerma did not exhibit any dependence on dose rate and energy, in contrast to tube voltage and HVL measurements.
Conclusion
The measurement values of the SSD fall within the acceptable range of uncertainty, highlighting its measurement accuracy and reliability. Furthermore, based on the characteristics elucidated by this study, valuable insights are provided concerning the assurance of appropriate measurement values in clinical settings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiological Technology, Gunma Prefectural College of Health Sciences, Gunma, Japan.
- Department of Medical Radiology, Faculty of Medical Technology, Teikyo University, Tokyo, Japan.
- Department of Radiological Sciences, Graduate School of Human Health Sciences, Tokyo Metropolitan University, Tokyo, Japan.
- Faculty of Medical Technology, Niigata University of Health and Welfare, Niigata, Japan.
ORCID-Profile: Sho Maruyama, Michiharu Sekimoto
Zitieren
APA
Maruyama, S., Saitou, H., Negishi, T., & Sekimoto, M. (2024). Evaluation of the measurement accuracy and uncertainty of a solid‐state detector under diagnostic x‐ray beam conditions. Journal of Applied Clinical Medical Physics, 25(9). https://doi.org/10.1002/acm2.14476
Vancouver
Maruyama S, Saitou H, Negishi T, Sekimoto M. Evaluation of the measurement accuracy and uncertainty of a solid‐state detector under diagnostic x‐ray beam conditions. Journal of Applied Clinical Medical Physics. 2024;25(9). doi:10.1002/acm2.14476
BibTeX
@article{Maruyama2024,
author = {Maruyama, Sho and Saitou, Hiroki and Negishi, Toru and Sekimoto, Michiharu},
title = {Evaluation of the measurement accuracy and uncertainty of a solid‐state detector under diagnostic x‐ray beam conditions},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
volume = {25},
number = {9},
doi = {10.1002/acm2.14476}
}
EUVIMED