Impact of proton PBS machine operating parameters on the effectiveness of layer rescanning for interplay effect mitigation in lung SBRT treatment
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14342 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Xiaoying Liang und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Mayo Clinic, Jacksonville, Florida, USA. Hauptthemen laut MEDLINE: Lung Neoplasms, Radiotherapy Dosage, Phantoms, Imaging, Radiosurgery. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Rescanning is a common technique used in proton pencil beam scanning to mitigate the interplay effect. Advances in machine operating parameters across different generations of particle therapy systems have led to improvements in beam delivery time (BDT). However, the potential impact of these improvements on the effectiveness of rescanning remains an underexplored area in the existing research.
Methods
We systematically investigated the impact of proton machine operating parameters on the effectiveness of layer rescanning in mitigating interplay effect during lung SBRT treatment, using the CIRS phantom. Focused on the Hitachi synchrotron particle therapy system, we explored machine operating parameters from our institution's current (2015) and upcoming systems (2025A and 2025B). Accumulated dynamic 4D dose were reconstructed to assess the interplay effect and layer rescanning effectiveness.
Results
Achieving target coverage and dose homogeneity within 2% deviation required 6, 6, and 20 times layer rescanning for the 2015, 2025A, and 2025B machine parameters, respectively. Beyond this point, further increasing the number of layer rescanning did not further improve the dose distribution. BDTs without rescanning were 50.4, 24.4, and 11.4 s for 2015, 2025A, and 2025B, respectively. However, after incorporating proper number of layer rescanning (six for 2015 and 2025A, 20 for 2025B), BDTs increased to 67.0, 39.6, and 42.3 s for 2015, 2025A, and 2025B machine parameters. Our data also demonstrated the potential problem of false negative and false positive if the randomness of the respiratory phase at which the beam is initiated is not considered in the evaluation of interplay effect.
Conclusion
The effectiveness of layer rescanning for mitigating interplay effect is affected by machine operating parameters. Therefore, past clinical experiences may not be applicable to modern machines.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Mayo Clinic, Jacksonville, Florida, USA.
- Department of Radiation Oncology, The First Affiliated Hospital of Zhengzhou University, Zhengzhou, China.
- Department of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA.
- Department of Radiation Oncology, Winship Cancer Institute, Emory University, Atlanta, USA.
ORCID-Profile: Xiaoying Liang, Jiajian Shen, Sridhar Yaddanapudi
Zitieren
APA
Liang, X., Liu, C., Shen, J., Flampouri, S., Park, J. C., Lu, B., Yaddanapudi, S., Tan, J., Furutani, K. M., & Beltran, C. J. (2024). Impact of proton PBS machine operating parameters on the effectiveness of layer rescanning for interplay effect mitigation in lung SBRT treatment. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14342
Vancouver
Liang X, Liu C, Shen J, Flampouri S, Park JC, Lu B, et al. Impact of proton PBS machine operating parameters on the effectiveness of layer rescanning for interplay effect mitigation in lung SBRT treatment. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14342
BibTeX
@article{Liang2024,
author = {Liang, Xiaoying and Liu, Chunbo and Shen, Jiajian and Flampouri, Stella and Park, Justin C and Lu, Bo and Yaddanapudi, Sridhar and Tan, Jun and Furutani, Keith M and Beltran, Chris J},
title = {Impact of proton PBS machine operating parameters on the effectiveness of layer rescanning for interplay effect mitigation in lung SBRT treatment},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14342}
}
EUVIMED