Characterization of the IDENTIFY TM surface scanning system for radiation therapy setup on a closed‐bore linac

Janita Dekker iD, Sander van Het Schip, Marion Essers, Mariska de Smet, Martijn Kusters, Willy de Kruijf

Journal of Applied Clinical Medical Physics · 2024-04 · 25 (4)

DOI: 10.1002/acm2.14326 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, online erschienen am 18.03.2024. Verfasst von Janita Dekker und 5 weiteren Autor:innen. Beteiligte Einrichtung: Institute Verbeeten, Tilburg, Netherlands. Hauptthemen laut MEDLINE: Radiotherapy, Image-Guided. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

In radiation therapy, surface guidance can be used for patient setup and intra‐fraction motion monitoring. The surface guided radiation therapy (SGRT) system from Varian Medical systems, IDENTIFY TM , consists of three pods, including cameras and a random pattern projector, mounted on the ceiling. The information captured by the cameras is used to make a reconstruction of the surface. The aim of the study was to assess the technical performance of this SGRT system on a closed‐bore linac.

Methods

Phantom measurements were performed to assess the accuracy, precision, reproducibility and temporal stability of the system, both in align and in load position. Translations of the phantoms in lateral, longitudinal, and vertical direction, and rotations around three axes (pitch, roll and yaw) were performed with an accurate, in‐house built, positioning stage. Different phantom geometries and different surface colors were used, and various ambient light intensities were tested.

Results

The accuracy of the IDENTIFY TM system at the closed‐bore linac was 0.07 mm and 0.07 degrees at load position, and 0.06 mm and 0.01 degrees at align position for the white head phantom. The precision was 0.02 mm and 0.02 degrees in load position, and 0.01 mm and 0.02 degrees in align position. The accuracy for the Penta‐Guide phantom was comparable to the white head phantom, with 0.06 mm and 0.01 degrees in align position. The system was slightly less accurate for translations of the CIRS lung phantom in align position (0.20 mm, 0.05 degrees). Reproducibility measurements showed a variation of 0.02 mm in load position. Regarding the temporal stability, the maximum drift over 30 min was 0.33 mm and 0.02 degrees in load position. No effect of ambient light level on the accuracy of the IDENTIFY TM system was observed. Regarding different surface colors, the accuracy of the system for a black phantom was slightly worse compared to a white surface, but not clinical relevant.

Conclusion

The IDENTIFY TM system can adequately be used for motion monitoring on a closed‐bore linac with submillimeter accuracy. The results of the performed measurements meet the clinical requirements described in the guidelines of the AAPM and the ESTRO.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy, Image-GuidedHumansReproducibility of ResultsPhantoms, ImagingPatient PositioningRadiotherapy Planning, Computer-Assisted

Stichwörter der Autor:innen

accuracyphantom measurementssurface guided radiation therapytechnical performance

Autor:innen und Institutionen

  • Institute Verbeeten, Tilburg, Netherlands.
  • Radboud University Medical Center, Nijmegen, Netherlands.

ORCID-Profile: Janita Dekker

Zitieren

APA

Dekker, J., van Het Schip, S., Essers, M., de Smet, M., Kusters, M., & de Kruijf, W. (2024). Characterization of the IDENTIFY TM surface scanning system for radiation therapy setup on a closed‐bore linac. Journal of Applied Clinical Medical Physics, 25(4). https://doi.org/10.1002/acm2.14326

Vancouver

Dekker J, van Het Schip S, Essers M, de Smet M, Kusters M, de Kruijf W. Characterization of the IDENTIFY TM surface scanning system for radiation therapy setup on a closed‐bore linac. Journal of Applied Clinical Medical Physics. 2024;25(4). doi:10.1002/acm2.14326

BibTeX
@article{Dekker2024,
  author = {Dekker, Janita and van Het Schip, Sander and Essers, Marion and de Smet, Mariska and Kusters, Martijn and de Kruijf, Willy},
  title = {Characterization of the IDENTIFY TM surface scanning system for radiation therapy setup on a closed‐bore linac},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  volume = {25},
  number = {4},
  doi = {10.1002/acm2.14326}
}
Kostenlos starten →