Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Kareem Rayn und 6 weiteren Autor:innen. Beteiligte Einrichtung: Varian Medical Affairs, Palo Alto, California, USA. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Radiotherapy, Intensity-Modulated, Radiotherapy Dosage, Thoracic Wall. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose/Objective Field size limitations on Halcyon and Ethos treatment machines largely preclude use of the conventional monoisocentric three‐field technique for breast/chest wall and regional lymph nodes. We present an alternative, IMRT‐based planning approach that facilitates treatment on Halcyon and Ethos while preserving plan quality. Materials/Methods Eight breast and regional node cases (four left‐sided, four right‐sided) were planned for an Ethos machine using a 15–17 field IMRT technique. Institutional plan quality metrics for CTV and PTV coverage and OAR sparing were assessed. Five plans (four right‐sided, one left‐sided) were also planned using a hybrid 3D multisocenter technique. CTV coverage and OAR sparing were compared to the IMRT plans. Eclipse scripting tools were developed to aid in beam placement and plan evaluation through a set of dosimetric scorecards, and both are shared publicly.
Results
On average, the IMRT plans achieved breast CTV and PTV coverage at 50 Gy of 97.9% and 95.7%, respectively. Supraclavicular CTV and PTV coverages at 45 Gy were 100% and 95.5%. Axillary lymph node CTV and PTV coverages at 45 Gy were 100% and 97.1%, and IMN CTV coverage at 45 Gy was 99.2%. Mean ipsilateral lung V20 Gy was 19.3%, and average mean heart dose was 1.6 Gy for right‐sided cases and 3.0 Gy for left‐sided. In comparison to the hybrid 3D plans, IMRT plans achieved higher breast and supraclavicular CTV coverage (99.9% vs. 98.6% and 99.9% vs. 93.4%), higher IMN coverage (99.6% vs. 78.2%), and lower ipsilateral lung V20 Gy (19.6% vs. 28.2%).
Conclusion
Institutional plan quality benchmarks were achieved for all eight cases using the IMRT‐based planning approach. The IMRT‐based planning approach offered superior conformity and OAR sparing than a competing hybrid 3D approach.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Varian Medical Affairs, Palo Alto, California, USA.
- Department of Radiation Oncology, University of California, Los Angeles, California, USA.
ORCID-Profile: Klea Hoxha, Dylan P O'Connell
Zitieren
APA
Rayn, K., Clark, R., Hoxha, K., Magliari, A., Neylon, J., Xiang, M. H., & O'Connell, D. P. (2024). An IMRT planning technique for treating whole breast or chest wall with regional lymph nodes on Halcyon and Ethos. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14295
Vancouver
Rayn K, Clark R, Hoxha K, Magliari A, Neylon J, Xiang MH, et al. An IMRT planning technique for treating whole breast or chest wall with regional lymph nodes on Halcyon and Ethos. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14295
BibTeX
@article{Rayn2024,
author = {Rayn, Kareem and Clark, Ryan and Hoxha, Klea and Magliari, Anthony and Neylon, Jack and Xiang, Michael H and O'Connell, Dylan P},
title = {An IMRT planning technique for treating whole breast or chest wall with regional lymph nodes on Halcyon and Ethos},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14295}
}
EUVIMED