Total scalp irradiation: A study comparing multiple types of bolus and VMAT optimization techniques
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14260 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Tanisha M Davis und 8 weiteren Autor:innen. Beteiligte Einrichtung: Medical Dosimetry Program, Southern Illinois University, Carbondale, Illinois, USA. Hauptthemen laut MEDLINE: Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NIH HHS, NIBIB NIH HHS, NCI NIH HHS.
Abstract (Originalsprache)
Purpose
To investigate bolus design and VMAT optimization settings for total scalp irradiation.
Methods
Three silicone bolus designs (flat, hat, and custom) from .decimal were evaluated for adherence to five anthropomorphic head phantoms. Flat bolus was cut from a silicone sheet. Generic hat bolus resembles an elongated swim cap while custom bolus is manufactured by injecting silicone into a 3D printed mold. Bolus placement time was recorded. Air gaps between bolus and scalp were quantified on CT images. The dosimetric effect of air gaps on target coverage was evaluated in a treatment planning study where the scalp was planned to 60 Gy in 30 fractions. A noncoplanar VMAT technique based on gEUD penalties was investigated that explored the full range of gEUD alpha values to determine which settings achieve sufficient target coverage while minimizing brain dose. ANOVA and the t ‐test were used to evaluate statistically significant differences (threshold = 0.05).
Results
The flat bolus took 32 ± 5.9 min to construct and place, which was significantly longer ( p < 0.001) compared with 0.67 ± 0.2 min for the generic hat bolus or 0.53 ± 0.10 min for the custom bolus. The air gap volumes were 38 ± 9.3 cc, 32 ± 14 cc, and 17 ± 7.0 cc for the flat, hat, and custom boluses, respectively. While the air gap differences between the flat and custom boluses were significant ( p = 0.011), there were no significant dosimetric differences in PTV coverage at V57Gy or V60Gy. In the VMAT optimization study, a gEUD alpha of 2 was found to minimize the mean brain dose.
Conclusions
Two challenging aspects of total scalp irradiation were investigated: bolus design and plan optimization. Results from this study show opportunities to shorten bolus fabrication time during simulation and create high quality treatment plans using a straightforward VMAT template with simple optimization settings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Medical Dosimetry Program, Southern Illinois University, Carbondale, Illinois, USA.
- Department of Radiation Oncology, Emory University, Atlanta, Georgia, USA.
ORCID-Profile: Kirk Luca, Xiaofeng Yang, Jiahan Zhang
Förderung
- NIH HHS
- NIBIB NIH HHS
- NCI NIH HHS
- Winship Cancer Institute
- National Institutes of Health
Zitieren
APA
Davis, T. M., Luca, K., Sudmeier, L. J., Buchwald, Z. S., Khan, M. K., Yang, X., Schreibmann, E., Zhang, J., & Roper, J. (2024). Total scalp irradiation: A study comparing multiple types of bolus and VMAT optimization techniques. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14260
Vancouver
Davis TM, Luca K, Sudmeier LJ, Buchwald ZS, Khan MK, Yang X, et al. Total scalp irradiation: A study comparing multiple types of bolus and VMAT optimization techniques. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14260
BibTeX
@article{Davis2024,
author = {Davis, Tanisha M and Luca, Kirk and Sudmeier, Lisa J and Buchwald, Zachary S and Khan, Mohammad K and Yang, Xiaofeng and Schreibmann, Eduard and Zhang, Jiahan and Roper, Justin},
title = {Total scalp irradiation: A study comparing multiple types of bolus and VMAT optimization techniques},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14260}
}
EUVIMED