Detecting outliers beyond tolerance limits derived from statistical process control in patient‐specific quality assurance

Hong Qi Tan iD, Kah Seng Lew, Yun Ming Wong, Wen Chuan Chong, Calvin Wei Yang Koh, Clifford Ghee Ann Chua, Ping Lin Yeap, Khong Wei Ang, James Cheow Lei Lee, Sung Yong Park

Journal of Applied Clinical Medical Physics · 2024

DOI: 10.1002/acm2.14154 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Hong Qi Tan und 9 weiteren Autor:innen. Beteiligte Einrichtung: Division of Radiation Oncology, National Cancer Centre Singapore, Singapore, Singapore. Hauptthemen laut MEDLINE: Radiosurgery, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 60 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Academic Program Goh Foundation Proton Research Programme.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Tolerance limit is defined on pre‐treatment patient specific quality assurance results to identify “out of the norm” dose discrepancy in plan. An out‐of‐tolerance plan during measurement can often cause treatment delays especially if replanning is required. In this study, we aim to develop an outlier detection model to identify out‐of‐tolerance plan early during treatment planning phase to mitigate the above‐mentioned risks.

Methods

Patient‐specific quality assurance results with portal dosimetry for stereotactic body radiotherapy measured between January 2020 and December 2021 were used in this study. Data were divided into thorax and pelvis sites and gamma passing rates were recorded using 2%/2 mm, 2%/1 mm, and 1%/1 mm gamma criteria. Statistical process control method was used to determine six different site and criterion‐specific tolerance and action limits. Using only the inliers identified with our determined tolerance limits, we trained three different outlier detection models using the plan complexity metrics extracted from each treatment field—robust covariance, isolation forest, and one class support vector machine. The hyperparameters were optimized using the F1‐score calculated from both the inliers and validation outliers’ data.

Results

308 pelvis and 200 thorax fields were used in this study. The tolerance (action) limits for 2%/2 mm, 2%/1 mm, and 1%/1 mm gamma criteria in the pelvis site are 99.1% (98.1%), 95.8% (91.1%), and 91.7% (86.1%), respectively. The tolerance (action) limits in the thorax site are 99.0% (98.7%), 97.0% (96.2%), and 91.5% (87.2%). One class support vector machine performs the best among all the algorithms. The best performing model in the thorax (pelvis) site achieves a precision of 0.56 (0.54), recall of 1.0 (1.0), and F1‐score of 0.72 (0.70) when using the 2%/2 mm (2%/1 mm) criterion.

Conclusion

The model will help the planner to identify an out‐of‐tolerance plan early so that they can refine the plan further during the planning stage without risking late discovery during measurement.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryRadiotherapy, Intensity-ModulatedHumansRadiotherapy Planning, Computer-AssistedRadiotherapy DosageAlgorithmsPelvisRadiometryQuality Assurance, Health Care

Stichwörter der Autor:innen

outlier detectionpatient-specific quality assuranceportal dosimetrystatistical process control

Autor:innen und Institutionen

  • Division of Radiation Oncology, National Cancer Centre Singapore, Singapore, Singapore.
  • Oncology Academic Clinical Programme, Duke-NUS Medical School, Singapore, Singapore.
  • Division of Physics and Applied Physics, Nanyang Technological University, Singapore, Singapore.

ORCID-Profile: Hong Qi Tan

Förderung

  • Academic Program Goh Foundation Proton Research Programme

Zitieren

APA

Tan, H. Q., Lew, K. S., Wong, Y. M., Chong, W. C., Koh, C. W. Y., Chua, C. G. A., Yeap, P. L., Ang, K. W., Lee, J. C. L., & Park, S. Y. (2024). Detecting outliers beyond tolerance limits derived from statistical process control in patient‐specific quality assurance. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14154

Vancouver

Tan HQ, Lew KS, Wong YM, Chong WC, Koh CWY, Chua CGA, et al. Detecting outliers beyond tolerance limits derived from statistical process control in patient‐specific quality assurance. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14154

BibTeX
@article{Tan2024,
  author = {Tan, Hong Qi and Lew, Kah Seng and Wong, Yun Ming and Chong, Wen Chuan and Koh, Calvin Wei Yang and Chua, Clifford Ghee Ann and Yeap, Ping Lin and Ang, Khong Wei and Lee, James Cheow Lei and Park, Sung Yong},
  title = {Detecting outliers beyond tolerance limits derived from statistical process control in patient‐specific quality assurance},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2024},
  doi = {10.1002/acm2.14154}
}
Infokarte | EUVIMED

Artikeldaten derzeit nicht verfügbar.

Recherche