EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Dosimetric verification of four dose calculation algorithms for spine stereotactic body radiotherapy

Hideaki Hirashima, Mitsuhiro Nakamura, Kiyonao Nakamura, Yukinori Matsuo, Takashi Mizowaki

Journal of Radiation Research · 2024

DOI: 10.1093/jrr/rrad086 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Radiation Research, 2024. Verfasst von Hideaki Hirashima und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology and Image-Applied Therapy, Graduate School of Medicine, Kyoto University, 54 Kawahara-cho, Shogoin, Sakyo-ku, Kyoto …. Hauptthemen laut MEDLINE: Radiosurgery, Radiotherapy, Intensity-Modulated. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: JSPS KAKENHI.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The applications of Type B [anisotropic analytical algorithm (AAA) and collapsed cone (CC)] and Type C [Acuros XB (AXB) and photon Monte Carlo (PMC)] dose calculation algorithms in spine stereotactic body radiotherapy (SBRT) were evaluated. Water- and bone-equivalent phantoms were combined to evaluate the percentage depth dose and dose profile. Subsequently, 48 consecutive patients with clinical spine SBRT plans were evaluated. All treatment plans were created using AXB in Eclipse. The prescription dose was 24 Gy in two fractions at a 10 MV FFF on TrueBeam. The doses were then recalculated with AAA, CC and PMC while maintaining the AXB-calculated monitor units and beam arrangement. The dose index values obtained using the four dose calculation algorithms were then compared. The AXB and PMC dose distributions agreed with the bone-equivalent phantom measurements (within ±2.0%); the AAA and CC values were higher than those in the bone-equivalent phantom region. For the spine SBRT plans, PMC, AAA and CC were overestimated compared with AXB in terms of the near minimum and maximum doses of the target and organ at risk, respectively; the mean dose difference was within 4.2%, which is equivalent with within 1 Gy. The phantom study showed that the results from AXB and PMC agreed with the measurements within ±2.0%. However, the mean dose difference ranged from 0.5 to 1 Gy in the spine SBRT planning study when the dose calculation algorithms changed. Users should incorporate a clinical introduction that includes an awareness of these differences.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryRadiotherapy, Intensity-ModulatedHumansRadiotherapy DosageRadiotherapy Planning, Computer-AssistedRadiometryAlgorithmsPhantoms, Imaging

Stichwörter der Autor:innen

dose calculation algorithmdosimetric verificationspine SBRT

Autor:innen und Institutionen

  • Department of Radiation Oncology and Image-Applied Therapy, Graduate School of Medicine, Kyoto University, 54 Kawahara-cho, Shogoin, Sakyo-ku, Kyoto 606-8507, Japan.
  • Department of Advanced Medical Physics, Graduate School of Medicine, Kyoto University, 53 Kawahara-cho, Shogoin, Sakyo-ku, Kyoto 606-8507, Japan.

Förderung

  • JSPS KAKENHI

Zitieren

APA

Hirashima, H., Nakamura, M., Nakamura, K., Matsuo, Y., & Mizowaki, T. (2024). Dosimetric verification of four dose calculation algorithms for spine stereotactic body radiotherapy. Journal of Radiation Research. https://doi.org/10.1093/jrr/rrad086

Vancouver

Hirashima H, Nakamura M, Nakamura K, Matsuo Y, Mizowaki T. Dosimetric verification of four dose calculation algorithms for spine stereotactic body radiotherapy. Journal of Radiation Research. 2024. doi:10.1093/jrr/rrad086

BibTeX
@article{Hirashima2024,
  author = {Hirashima, Hideaki and Nakamura, Mitsuhiro and Nakamura, Kiyonao and Matsuo, Yukinori and Mizowaki, Takashi},
  title = {Dosimetric verification of four dose calculation algorithms for spine stereotactic body radiotherapy},
  journal = {Journal of Radiation Research},
  year = {2024},
  doi = {10.1093/jrr/rrad086}
}
Kostenlos starten →