How much rotational error is clinically acceptable for single‐isocenter/two‐lesion lung SBRT treatment on halcyon ring delivery system (RDS)?

Damodar Pokhrel, Richard Mallory, Mark E Bernard, Mahesh Kudrimoti

Journal of Applied Clinical Medical Physics · 2023

DOI: 10.1002/acm2.14068 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2023. Verfasst von Damodar Pokhrel und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Medicine, University of Kentucky, Lexington, Kentucky, USA. Hauptthemen laut MEDLINE: Radiosurgery, Radiotherapy, Intensity-Modulated, Lung Neoplasms. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

SBRT treatment of two separate lung lesions via single‐isocenter/multi‐target (SIMT) plan on Halcyon RDS could improve patient comfort, compliance, patient throughput, and clinic efficiency. However, aligning two separate lung lesions synchronously via a single pre‐treatment CBCT scan on Halcyon can be difficult due to rotational patient setup errors. Thus, to quantify the dosimetric impact, we simulated loss of target(s) coverage due to small, yet clinically observable rotational patient setup errors on Halcyon for SIMT treatments.

Methods

Seventeen previously treated 4D‐CT based SIMT lung SBRT patients with two separate lesions (total 34 lesions, 50 Gy in five fractions to each lesion) on TrueBeam (6MV‐FFF) were re‐planned on Halcyon (6MV‐FFF) using a similar arc geometry (except couch rotation), dose engine (AcurosXB algorithm), and treatment planning objectives. Rotational patient setup errors of [± 0.5⁰ to ± 3.0⁰] on Halcyon were simulated via Velocity registration software in all three rotation axes and recalculated dose distributions in Eclipse treatment planning system. Dosimetric impact of rotational errors was evaluated for target coverage and organs at risk (OAR).

Results

Average PTV volume and distance to isocenter were 23.7 cc and 6.1 cm. Average change in Paddick's conformity indexes were less than −5%, −10%, and −15% for 1°, 2°, and 3°, respectively for yaw, roll, and pitch rotation directions. Maximum drop off of PTV(D100%) coverage for 2° rotation was −2.0% (yaw), −2.2% (roll), and −2.5% (pitch). With ±1° rotational error, no PTV(D100%) loss was found. Due to anatomical complexity: irregular and highly variable tumor sizes and locations, highly heterogenous dose distribution, and steep dose gradient, no trend for loss of target(s) coverage as a function of distance to isocenter and PTV size was found. Change in maximum dose to OAR were acceptable per NRG‐BR001 within ±1.0° rotation, but were up to 5 Gy higher to heart with 2° in the pitch rotation axis.

Conclusion

Our clinically realistic simulation results show that rotational patient setup errors up to 1.0° in any rotation axis could be acceptable for selected two separate lung lesions SBRT patients on Halcyon. Multivariable data analysis in large cohort is ongoing to fully characterize Halcyon RDS for synchronous SIMT lung SBRT.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryRadiotherapy, Intensity-ModulatedLung NeoplasmsHumansRadiotherapy Planning, Computer-AssistedRadiotherapy DosageRetrospective StudiesLung

Stichwörter der Autor:innen

Halcyon RDSSIMTpatient set up errorssynchronous Lung SBRTtarget coverage loss

Autor:innen und Institutionen

  • Department of Radiation Medicine, University of Kentucky, Lexington, Kentucky, USA.

Zitieren

APA

Pokhrel, D., Mallory, R., Bernard, M. E., & Kudrimoti, M. (2023). How much rotational error is clinically acceptable for single‐isocenter/two‐lesion lung SBRT treatment on halcyon ring delivery system (RDS)?. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14068

Vancouver

Pokhrel D, Mallory R, Bernard ME, Kudrimoti M. How much rotational error is clinically acceptable for single‐isocenter/two‐lesion lung SBRT treatment on halcyon ring delivery system (RDS)?. Journal of Applied Clinical Medical Physics. 2023. doi:10.1002/acm2.14068

BibTeX
@article{Pokhrel2023,
  author = {Pokhrel, Damodar and Mallory, Richard and Bernard, Mark E and Kudrimoti, Mahesh},
  title = {How much rotational error is clinically acceptable for single‐isocenter/two‐lesion lung SBRT treatment on halcyon ring delivery system (RDS)?},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2023},
  doi = {10.1002/acm2.14068}
}
Kostenlos starten →