Influence of air mapping errors on the dosimetric accuracy of prostate CBCT‐guided online adaptive radiation therapy

Olga M Dona Lemus iD, Sean Tanny, Michael Cummings, Matthew Webster, Joshua Wancura, Hyunuk Jung, Yuwei Zhou, Jihyung Yoon, Matthew Pacella, Dandan Zheng

Journal of Applied Clinical Medical Physics · 2023

DOI: 10.1002/acm2.14057 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2023. Verfasst von Olga M Dona Lemus und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Rochester Medical Center, New York, New York, USA. Hauptthemen laut MEDLINE: Prostate, Spiral Cone-Beam Computed Tomography. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

CBCT‐guided online adaptive radiotherapy (oART) plans presently utilize daily synthetic CTs (sCT) that are automatically generated using deformable registration algorithms. These algorithms may have poor performance at reproducing variable volumes of gas present during treatment. Therefore, we have analyzed the air mapping error between the daily CBCTs and the corresponding sCT and explored its dosimetric effect on oART plan calculation.

Methods

Abdominopelvic air volume was contoured on both the daily CBCT images and the corresponding synthetic images for 207 online adaptive pelvic treatments. Air mapping errors were tracked over all fractions. For two case studies representing worst case scenarios, dosimetric effects of air mapping errors were corrected in the sCT images using the daily CBCT air contours, then recalculating dose. Dose volume histogram statistics and 3D gamma passing rates were used to compare the original and air‐corrected sCT‐based dose calculations.

Results

All analyzed patients showed observable air pocket contour differences between the sCT and the CBCT images. The largest air volume difference observed in daily CBCT images for a given patient was 276.3 cc, a difference of more than 386% compared to the sCT. For the two case studies, the largest observed change in DVH metrics was a 2.6% reduction in minimum PTV dose, with all other metrics varying by less than 1.5%. 3D gamma passing rates using 1%/1 mm criteria were above 90% when comparing the uncorrected and corrected dose distributions.

Conclusion

Current CBCT‐based oART workflow can lead to inaccuracies in the mapping of abdominopelvic air pockets from daily CBCT to the sCT images used for the optimization and calculation of the adaptive plan. Despite the large observed mapping errors, the dosimetric effects of such differences on the accuracy of the adapted plan dose calculation are unlikely to cause differences greater than 3% for prostate treatments.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ProstateSpiral Cone-Beam Computed TomographyMaleHumansRadiotherapy DosageRadiotherapy Planning, Computer-AssistedCone-Beam Computed Tomography

Stichwörter der Autor:innen

adaptive therapyair mapping errorsprostate cancersynthetic CT

Autor:innen und Institutionen

  • Department of Radiation Oncology, University of Rochester Medical Center, New York, New York, USA.

ORCID-Profile: Olga M Dona Lemus

Zitieren

APA

Lemus, O. M. D., Tanny, S., Cummings, M., Webster, M., Wancura, J., Jung, H., Zhou, Y., Yoon, J., Pacella, M., & Zheng, D. (2023). Influence of air mapping errors on the dosimetric accuracy of prostate CBCT‐guided online adaptive radiation therapy. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14057

Vancouver

Lemus OMD, Tanny S, Cummings M, Webster M, Wancura J, Jung H, et al. Influence of air mapping errors on the dosimetric accuracy of prostate CBCT‐guided online adaptive radiation therapy. Journal of Applied Clinical Medical Physics. 2023. doi:10.1002/acm2.14057

BibTeX
@article{Lemus2023,
  author = {Lemus, Olga M Dona and Tanny, Sean and Cummings, Michael and Webster, Matthew and Wancura, Joshua and Jung, Hyunuk and Zhou, Yuwei and Yoon, Jihyung and Pacella, Matthew and Zheng, Dandan},
  title = {Influence of air mapping errors on the dosimetric accuracy of prostate CBCT‐guided online adaptive radiation therapy},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2023},
  doi = {10.1002/acm2.14057}
}
Kostenlos starten →