Dosimetric evaluation of a spinal cord dose‐limiting 3D‐CRT technique for radiotherapy of spinal metastases
Journal of Applied Clinical Medical Physics · 2023
DOI: 10.1002/acm2.14042 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2023. Verfasst von Christian Cornelius Arnold und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiotherapy and Oncology, Goethe University Frankfurt, Frankfurt, Germany. Hauptthemen laut MEDLINE: Spinal Neoplasms, Radiotherapy, Conformal, Radiotherapy, Intensity-Modulated. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
To evaluate the possible advantages of a simple spinal cord (SC) dose‐limiting three‐dimensional conformal radiotherapy (3D‐CRT) technique in comparison to conventional two‐dimensional (2D) techniques and other 3D‐CRT techniques for spinal bone irradiation.
Methods
For 41 spinal target volumes, seven different techniques were evaluated, using a standard schedule of 30 Gy in 10 fractions. The SC dose‐limiting 3D‐CRT technique 1F2S‐18MV using a single posterior field (F) supplemented by two anterior segment fields (S) and 18‐MV photon beams was compared to two conventional 2D techniques (a single posterior field, PA , and two opposed anterior‐posterior fields, APPA ), three other 3D‐CRT techniques (a single posterior field supplemented by four segment fields, 1F4S ; two wedged fields, WD , and the SC dose‐limiting variant using 6 MV, 1F2S‐6MV ) along with the original clinically applied plans.
Results
1F2S‐18MV demonstrated notably better results for all target volume parameters compared to the conventional 2D techniques ( p < 0.001). Limitation of the SC dose was significantly superior with 1F2S‐18MV in comparison to PA and APPA (SC Dmean: 28.9 ± 0.4 vs. 30.1 ± 0.6 Gy and 30.1 ± 0.4 Gy; SC Dmax: 30.9 ± 0.7 vs. 32.5 ± 1.0 Gy and 31.8 ± 0.7 Gy; SC D1cm 3 : 30.1 ± 0.6 vs. 31.7 ± 0.9 Gy and 31.1 ± 0.6 Gy; p < 0.001). Likewise, lower mean SC doses with 1F2S‐18MV were observed in comparison to the more treatment time‐consuming 3D‐CRT techniques (1F4S, WD) and the original plans without relevant compromises on the dose homogeneity in the target volume and the dose exposure to the other OARs.
Conclusion
In treatment planning of spinal metastases, simple variants of 3D‐CRT‐techniques like 1F2S‐18MV can offer a significant dose limitation to the SC while providing a sufficient dose coverage of the target volume. Especially in patients with favorable life expectancy and potential need for re‐irradiation, such SC dose‐limiting 3D‐CRT techniques may be a reasonable approach.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiotherapy and Oncology, Goethe University Frankfurt, Frankfurt, Germany.
- Department of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
ORCID-Profile: Christian Cornelius Arnold, Klaus Bratengeier
Zitieren
APA
Arnold, C. C., Toussaint, A., Mantel, F., Flentje, M., & Bratengeier, K. (2023). Dosimetric evaluation of a spinal cord dose‐limiting 3D‐CRT technique for radiotherapy of spinal metastases. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14042
Vancouver
Arnold CC, Toussaint A, Mantel F, Flentje M, Bratengeier K. Dosimetric evaluation of a spinal cord dose‐limiting 3D‐CRT technique for radiotherapy of spinal metastases. Journal of Applied Clinical Medical Physics. 2023. doi:10.1002/acm2.14042
BibTeX
@article{Arnold2023,
author = {Arnold, Christian Cornelius and Toussaint, André and Mantel, Frederick and Flentje, Michael and Bratengeier, Klaus},
title = {Dosimetric evaluation of a spinal cord dose‐limiting 3D‐CRT technique for radiotherapy of spinal metastases},
journal = {Journal of Applied Clinical Medical Physics},
year = {2023},
doi = {10.1002/acm2.14042}
}
EUVIMED