Effect of scattered megavoltage x‐rays on markerless tumor tracking using dual energy kilovoltage imaging
Journal of Applied Clinical Medical Physics · 2023
DOI: 10.1002/acm2.13993 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2023. Verfasst von Mandeep Kaur und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Stritch School of Medicine, Cardinal Bernadin Cancer Center, Loyola University Chicago, Maywood, Illinois, USA. Hauptthemen laut MEDLINE: Lung Neoplasms. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS, NIH HHS, National Institutes of Health.
Abstract (Originalsprache)
Purpose
To determine the effect of megavoltage (MV) scatter on the accuracy of markerless tumor tracking (MTT) for lung tumors using dual energy (DE) imaging and to consider a post‐processing technique to mitigate the effects of MV scatter on DE‐MTT.
Methods
A Varian TrueBeam linac was used to acquire a series of interleaved 60/120 kVp images of a motion phantom with simulated tumors (10 and 15 mm diameter). Two sets of consecutive high/low energy projections were acquired, with and without MV beam delivery. The MV field sizes (FS) ranged from 2 × 2 cm 2 –6 × 6 cm 2 in steps of 1 × 1 cm 2 . Weighted logarithmic subtraction was performed on sequential images to produce soft‐tissue images for kV only (DE kV ) and kV with MV beam on (DE kV+MV ). Wavelet and fast Fourier transformation filtering (wavelet‐FFT) was used to remove stripe noise introduced by MV scatter in the DE images (). A template‐based matching algorithm was then used to track the target on DE kV, DE kV+MV , and images. Tracking accuracy was evaluated using the tracking success rate (TSR) and mean absolute error (MAE).
Results
For the 10 and 15 mm targets, the TSR for DE kV images was 98.7% and 100%, and MAE was 0.53 and 0.42 mm, respectively. For the 10 mm target, the TSR, including the effects of MV scatter, ranged from 86.5% (2 × 2 cm 2 ) to 69.4% (6 × 6 cm 2 ), while the MAE ranged from 2.05 mm to 4.04 mm. The application of wavelet‐FFT algorithm to remove stripe noise () resulted in TSR values of 96.9% (2 × 2 cm 2 ) to 93.4% (6 × 6 cm 2 ) and subsequent MAE values were 0.89 mm to 1.37 mm. Similar trends were observed for the 15 mm target.
Conclusion
MV scatter significantly impacts the tracking accuracy of lung tumors using DE images. Wavelet‐FFT filtering can improve the accuracy of DE‐MTT during treatment.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Stritch School of Medicine, Cardinal Bernadin Cancer Center, Loyola University Chicago, Maywood, Illinois, USA.
- Varian Medical Systems, Palo Alto, California, USA.
- Department of Radiation Oncology, Loyola University Medical Center, Maywood, Illinois, USA.
Förderung
- NCI NIH HHS
- NIH HHS
- National Institutes of Health
Zitieren
APA
Kaur, M., Luce, J., Lehmann, M., Morf, D., Zhu, L., Kang, H., Walczak, M., Harkenrider, M. M., & Roeske, J. C. (2023). Effect of scattered megavoltage x‐rays on markerless tumor tracking using dual energy kilovoltage imaging. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.13993
Vancouver
Kaur M, Luce J, Lehmann M, Morf D, Zhu L, Kang H, et al. Effect of scattered megavoltage x‐rays on markerless tumor tracking using dual energy kilovoltage imaging. Journal of Applied Clinical Medical Physics. 2023. doi:10.1002/acm2.13993
BibTeX
@article{Kaur2023,
author = {Kaur, Mandeep and Luce, Jason and Lehmann, Mathias and Morf, Daniel and Zhu, Liangjia and Kang, Hyejoo and Walczak, Michal and Harkenrider, Matthew M and Roeske, John C},
title = {Effect of scattered megavoltage x‐rays on markerless tumor tracking using dual energy kilovoltage imaging},
journal = {Journal of Applied Clinical Medical Physics},
year = {2023},
doi = {10.1002/acm2.13993}
}
EUVIMED