Dosimetric accuracy of the Convolution algorithm for Leksell Gamma Plan radiosurgery treatment planning: Evaluation in the presence of clinically relevant inhomogeneities
Journal of Applied Clinical Medical Physics · 2023
DOI: 10.1002/acm2.13903 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2023. Verfasst von Evaggelos Pantelis und 8 weiteren Autor:innen. Beteiligte Einrichtung: Medical Physics Laboratory, Medical School, National and Kapodistrian University of Athens, Athens, Greece. Hauptthemen laut MEDLINE: Radiosurgery. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: This work was supported by Elekta AB.
Abstract (Originalsprache)
Purpose
The Leksell Gamma Plan Convolution algorithm (LGP‐Convolution) has not been widely adopted. This mainly stems from the higher calculated beam‐on times relative to the standard ray tracing‐based LGP‐TMR10 dose calculation algorithm. This study aims to evaluate the accuracy of the LGP‐Convolution in scenarios where the treated lesions are in the vicinity of or encompassed by bone and/or air inhomogeneities.
Methods
The solid water dosimetry phantom provided by the vendor was modified to include bone and air inhomogeneities. Two treatment planning scenarios were investigated involving a single shot and multiple shots, respectively. Treatment planning and dose prescription were performed using the LGP‐Convolution algorithm. Triple channel film dosimetry was performed using GafChromic EBT3 films calibrated in terms of absorbed dose to water in a 60 Co beam. Monte Carlo (MC) simulation dosimetry was also performed in the inhomogeneous experimental geometry using the EGSnrc MC platform and a previously validated sector‐based phase‐space source model. MC simulations were also employed to determine correction factors required for converting EBT3 measurements at points within the bone and air inhomogeneities from dose‐to‐water values to the corresponding dose to medium values. Results and Conclusions EBT3 dose to medium correction factors ranged with field size (4, 8, or 16 mm) within 0.941–0.946 for bone and 0.745–0.749 for air inhomogeneities. An excellent agreement was found between the LGP‐Convolution calculations with corresponding EBT3 and MC dose to medium results at all measurement points, except those located inside the air inhomogeneity. The latter is of no clinical importance and excluding them yielded gamma index passing rates of nearly 100% for 3% local dose difference and 1 mm distance‐to‐agreement criteria. The excellent agreement observed between LGP‐Convolution calculations and film as well as MC results of dose to medium indicates that the latter is the quantity reported by the LGP‐Convolution.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Medical Physics Laboratory, Medical School, National and Kapodistrian University of Athens, Athens, Greece.
- Medical Genesis Care Centre for Radiotherapy, Cromwell Hospital, London, UK.
- Elekta Instrument AB, Stockholm, Sweden.
- Departments of Medical Physics and Gamma Knife, Hygeia Hospital, Athens, Greece.
Förderung
- This work was supported by Elekta AB
Zitieren
APA
Pantelis, E., Logothetis, A., Zoros, E., Pappas, E. P., Papagiannis, P., Paddick, I., Nordström, H., Kollias, G., & Karaiskos, P. (2023). Dosimetric accuracy of the Convolution algorithm for Leksell Gamma Plan radiosurgery treatment planning: Evaluation in the presence of clinically relevant inhomogeneities. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.13903
Vancouver
Pantelis E, Logothetis A, Zoros E, Pappas EP, Papagiannis P, Paddick I, et al. Dosimetric accuracy of the Convolution algorithm for Leksell Gamma Plan radiosurgery treatment planning: Evaluation in the presence of clinically relevant inhomogeneities. Journal of Applied Clinical Medical Physics. 2023. doi:10.1002/acm2.13903
BibTeX
@article{Pantelis2023,
author = {Pantelis, Evaggelos and Logothetis, Andreas and Zoros, Emmanouil and Pappas, Eleftherios P and Papagiannis, Panagiotis and Paddick, Ian and Nordström, Håkan and Kollias, George and Karaiskos, Pantelis},
title = {Dosimetric accuracy of the Convolution algorithm for Leksell Gamma Plan radiosurgery treatment planning: Evaluation in the presence of clinically relevant inhomogeneities},
journal = {Journal of Applied Clinical Medical Physics},
year = {2023},
doi = {10.1002/acm2.13903}
}
EUVIMED