EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Transitioning from a COMS‐based plaque brachytherapy program to using eye physics plaques and plaque simulator treatment planning system: A single institutional experience

Sheridan G Meltsner, Anna Rodrigues, Miguel A Materin, David G Kirsch, Oana Craciunescu

Journal of Applied Clinical Medical Physics · 2023

DOI: 10.1002/acm2.13902 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2023. Verfasst von Sheridan G Meltsner und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Duke University, Durham, North Carolina, USA. Hauptthemen laut MEDLINE: Brachytherapy, Eye Neoplasms, Melanoma. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 45 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The aim of this work is to describe the implementation and commissioning of a plaque brachytherapy program using Eye Physics eye plaques and Plaque Simulator treatment planning system based on the experience of one institution with an established COMS‐based plaque program. Although commissioning recommendations are available in official task groups publications such as TG‐129 and TG‐221, we found that there was a lack of published experiences with the specific details of such a transition and the practical application of the commissioning guidelines. The specific issues addressed in this paper include discussing the lack of FDA approval of the Eye Physics plaques and Plaque Simulator treatment planning system, the commissioning of the plaques and treatment planning system including considerations of the heterogeneity corrected calculations, and the implementation of a second check using an FDA‐approved treatment planning system. We have also discussed the use of rental plaques, the analysis of plans using dose histograms, and the development of a quality management program. By sharing our experiences with the commissioning of this program this document will assist other institutions with the same task and act as a supplement to the recommendations in the recently published TG‐221.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BrachytherapyEye NeoplasmsMelanomaHumansRadiotherapy DosageIodine RadioisotopesMonte Carlo MethodRadiotherapy Planning, Computer-Assisted

Stichwörter der Autor:innen

Eye Physics eye plaquesPlaque Simulator treatment planning systemcommissioningocular brachytherapyplaque brachytherapy

Substanzen

Iodine Radioisotopes

Autor:innen und Institutionen

  • Department of Radiation Oncology, Duke University, Durham, North Carolina, USA.
  • Departments of Ophthalmology, Duke University, Durham, North Carolina, USA.

Zitieren

APA

Meltsner, S. G., Rodrigues, A., Materin, M. A., Kirsch, D. G., & Craciunescu, O. (2023). Transitioning from a COMS‐based plaque brachytherapy program to using eye physics plaques and plaque simulator treatment planning system: A single institutional experience. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.13902

Vancouver

Meltsner SG, Rodrigues A, Materin MA, Kirsch DG, Craciunescu O. Transitioning from a COMS‐based plaque brachytherapy program to using eye physics plaques and plaque simulator treatment planning system: A single institutional experience. Journal of Applied Clinical Medical Physics. 2023. doi:10.1002/acm2.13902

BibTeX
@article{Meltsner2023,
  author = {Meltsner, Sheridan G and Rodrigues, Anna and Materin, Miguel A and Kirsch, David G and Craciunescu, Oana},
  title = {Transitioning from a COMS‐based plaque brachytherapy program to using eye physics plaques and plaque simulator treatment planning system: A single institutional experience},
  journal = {Journal of Applied Clinical Medical Physics},
  year = {2023},
  doi = {10.1002/acm2.13902}
}
Kostenlos starten →