Effect of Disposable Elevator Cap Duodenoscopes on Persistent Microbial Contamination and Technical Performance of Endoscopic Retrograde Cholangiopancreatography
JAMA Internal Medicine · 2023
DOI: 10.1001/jamainternmed.2022.6394 ↗Kurz gefasst
Fachartikel in der Zeitschrift JAMA Internal Medicine, 2023. Verfasst von Nauzer Forbes und 31 weiteren Autor:innen. Beteiligte Einrichtung: Division of Gastroenterology and Hepatology, Department of Medicine, University of Calgary, Calgary, Alberta, Canada. Hauptthemen laut MEDLINE: Duodenoscopes, Cholangiopancreatography, Endoscopic Retrograde. Die Arbeit wurde bisher 49-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: CIHR.
Abstract (Originalsprache)
Importance Infection transmission following endoscopic retrograde cholangiopancreatography (ERCP) can occur due to persistent contamination of duodenoscopes despite high-level disinfection to completely eliminate microorganisms on the instrument.
Objective
To determine (1) contamination rates after high-level disinfection and (2) technical performance of duodenoscopes with disposable elevator caps compared with those with standard designs. Design, Setting, and Participants In this parallel-arm multicenter randomized clinical trial at 2 tertiary ERCP centers in Canada, all patients 18 years and older and undergoing ERCP for any indication were eligible.
Intervention
The intervention was use of duodenoscopes with disposable elevator caps compared with duodenoscopes with a standard design. Main Outcomes and Measures Coprimary outcomes were persistent microbial contamination of the duodenoscope elevator or channel, defined as growth of at least 10 colony-forming units of any organism or any growth of gram-negative bacteria following high-level disinfection (superiority outcome), and technical success of ERCP according to a priori criteria (noninferiority outcome with an a priori noninferiority margin of 7%), assessed by blinded reviewers.
Results
From December 2019 to February 2022, 518 patients were enrolled (259 disposable elevator cap duodenoscopes, 259 standard duodenoscopes). Patients had a mean (SD) age of 60.7 (17.0) years and 258 (49.8%) were female. No significant differences were observed between study groups, including in ERCP difficulty. Persistent microbial contamination was detected in 11.2% (24 of 214) of standard duodenoscopes and 3.8% (8 of 208) of disposable elevator cap duodenoscopes ( P = .004), corresponding to a relative risk of 0.34 (95% CI, 0.16-0.75) and number needed to treat of 13.6 (95% CI, 8.1-42.7) to avoid persistent contamination. Technical success using the disposable cap scope was noninferior to that of the standard scope (94.6% vs 90.7%, P = .13). There were no differences between study groups in adverse events and other secondary outcomes. Conclusions and Relevance In this randomized clinical trial, disposable elevator cap duodenoscopes exhibited reduced contamination following high-level disinfection compared with standard scope designs, without affecting the technical performance and safety of ERCP. Trial Registration ClinicalTrials.gov Identifier: NCT04040504
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Division of Gastroenterology and Hepatology, Department of Medicine, University of Calgary, Calgary, Alberta, Canada.
- Department of Community Health Sciences, University of Calgary, Calgary, Alberta, Canada.
- Division of Gastroenterology and Hepatology, Medical University of South Carolina, Charleston.
- Department of Biological Science, University of Calgary, Calgary, Alberta, Canada.
- Department of Microbiology, Immunology and Infectious Diseases, University of Calgary, Calgary, Alberta, Canada.
- Division of Infectious Diseases, Department of Medicine, University of Calgary, Calgary, Alberta, Canada.
Förderung
- CIHR
Zitieren
APA
Forbes, N., Elmunzer, B. J., Allain, T., Parkins, M. D., Sheth, P. M., Waddell, B. J., Du, K., Douchant, K., Oladipo, O., Saleem, A., Cartwright, S., Chau, M., Howarth, M., McKay, J., Nashad, T., Ruan, Y., Bishay, K., Gonzalez-Moreno, E., Meng, Z. W., …, & Heitman, S. J. (2023). Effect of Disposable Elevator Cap Duodenoscopes on Persistent Microbial Contamination and Technical Performance of Endoscopic Retrograde Cholangiopancreatography. JAMA Internal Medicine. https://doi.org/10.1001/jamainternmed.2022.6394
Vancouver
Forbes N, Elmunzer BJ, Allain T, Parkins MD, Sheth PM, Waddell BJ, et al. Effect of Disposable Elevator Cap Duodenoscopes on Persistent Microbial Contamination and Technical Performance of Endoscopic Retrograde Cholangiopancreatography. JAMA Internal Medicine. 2023. doi:10.1001/jamainternmed.2022.6394
BibTeX
@article{Forbes2023,
author = {Forbes, Nauzer and Elmunzer, B Joseph and Allain, Thibault and Parkins, Michael D and Sheth, Prameet M and Waddell, Barbara J and Du, Kristine and Douchant, Katya and Oladipo, Olajumoke and Saleem, April and Cartwright, Shane and Chau, Millie and Howarth, Megan and McKay, Jackie and Nashad, Tamim and Ruan, Yibing and Bishay, Kirles and Gonzalez-Moreno, Emmanuel and Meng, Zhao Wu and Bass, Sydney and Bechara, Robert and Cole, Martin J and Jalink, Diederick W and Mohamed, Rachid and Turbide, Christian and Belletrutti, Paul J and Kayal, Ahmed and Kumar, Puja R and Hilsden, Robert J and Buret, André G and Hookey, Lawrence and Heitman, Steven J},
title = {Effect of Disposable Elevator Cap Duodenoscopes on Persistent Microbial Contamination and Technical Performance of Endoscopic Retrograde Cholangiopancreatography},
journal = {JAMA Internal Medicine},
year = {2023},
doi = {10.1001/jamainternmed.2022.6394}
}
EUVIMED