Endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscope reprocessing: Variables impacting contamination risk
Infection Control & Hospital Epidemiology · 2023
DOI: 10.1017/ice.2022.319 ↗Kurz gefasst
Beobachtungsstudie in der Zeitschrift Infection Control & Hospital Epidemiology, 2023. Verfasst von Ashley M Ayres und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Infection Prevention and Control, UPMC Presbyterian-Shadyside, Pittsburgh, Pennsylvania. Hauptthemen laut MEDLINE: Cholangiopancreatography, Endoscopic Retrograde, Disinfection. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 15 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-SA 4.0).
Abstract (Originalsprache)
Objective
To evaluate variables that affect risk of contamination for endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscopes.
Design
Observational, quality improvement study.
Setting
University medical center with a gastrointestinal endoscopy service performing ∼1,000 endoscopic retrograde cholangiopancreatography and ∼1,000 endoscopic ultrasound endoscope procedures annually.
Methods
Duodenoscope and linear echoendoscope sampling (from the elevator mechanism and instrument channel) was performed from June 2020 through September 2021. Operational changes during this period included standard reprocessing with high-level disinfection with ethylene oxide gas sterilization (HLD–ETO) was switched to double high-level disinfection (dHLD) (June 16, 2020–July 15, 2020), and duodenoscopes changed to disposable tip model (March 2021). The frequency of contamination for the co-primary outcomes were characterized by calculated risk ratios.
Results
The overall pathogenic contamination rate was 4.72% (6 of 127). Compared to duodenoscopes, linear echoendoscopes had a contamination risk ratio of 3.64 (95% confidence interval [CI], 0.69–19.1). Reprocessing using HLD-ETO was associated with a contamination risk ratio of 0.29 (95% CI, 0.06–1.54). Linear echoendoscopes undergoing dHLD had the highest risk of contamination (2 of 18, 11.1%), and duodenoscopes undergoing HLD-ETO and the lowest risk of contamination (0 of 53, 0%). Duodenoscopes with a disposable tip had a 0% contamination rate (0 of 27).
Conclusions
We did not detect a significant reduction in endoscope contamination using HLD-ETO versus dHLD reprocessing. Linear echoendoscopes have a risk of contamination similar to that of duodenoscopes. Disposable tips may reduce the risk of duodenoscope contamination.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Department of Infection Prevention and Control, UPMC Presbyterian-Shadyside, Pittsburgh, Pennsylvania.
- Department of Gastroenterology, Hepatology, and Nutrition, University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania.
- Division of Microbiology, Department of Pathology, University of Pittsburgh School of Medicine, Pittsburgh, Pennsylvania.
- Division of Infectious Diseases, Department of Medicine, University of Pittsburgh School of Medicine, Pittsburgh, Pennsylvania.
ORCID-Profile: Ashley M Ayres, A William Pasculle, Graham M Snyder
Zitieren
APA
Ayres, A. M., Wozniak, J., O'Neil, J., Stewart, K., Leger, J. S., Pasculle, A. W., Lewis, C., McGrath, K., Slivka, A., & Snyder, G. M. (2023). Endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscope reprocessing: Variables impacting contamination risk. Infection Control & Hospital Epidemiology. https://doi.org/10.1017/ice.2022.319
Vancouver
Ayres AM, Wozniak J, O'Neil J, Stewart K, Leger JS, Pasculle AW, et al. Endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscope reprocessing: Variables impacting contamination risk. Infection Control & Hospital Epidemiology. 2023. doi:10.1017/ice.2022.319
BibTeX
@article{Ayres2023,
author = {Ayres, Ashley M and Wozniak, Julia and O'Neil, Jose and Stewart, Kimberly and Leger, John St and Pasculle, A William and Lewis, Casey and McGrath, Kevin and Slivka, Adam and Snyder, Graham M},
title = {Endoscopic retrograde cholangiopancreatography and endoscopic ultrasound endoscope reprocessing: Variables impacting contamination risk},
journal = {Infection Control & Hospital Epidemiology},
year = {2023},
doi = {10.1017/ice.2022.319}
}
EUVIMED