Individualized Nonadaptive Image Guided Radiotherapy (IGRT) In Stage II-III Carcinoma Cervix With Daily Cone Beam Computed Tomography (CBCT): Set Up Error Calculation

Nilesh Mani, Harikesh Bahadur Singh, Poonam Gupta

IOSR Journal of Dental and Medical Sciences · 2026-07 · 25 (7) · 27-32

DOI: 10.9790/0853-2507012732 ↗

Kurz gefasst

Fachartikel in der Zeitschrift IOSR Journal of Dental and Medical Sciences, 2026-07. Verfasst von Nilesh Mani und 2 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The standard of care in stage II-III carcinoma cervix is radical concurrent chemo-radiotherapy followed by brachytherapy. This study aimed to give institutional set up error data calculation and comparative analysis of KV CBCT (Kilovoltage Cone-Beam Computed Tomography) and KV MV (Kilovoltage and Megavoltage) Imaging CBCT. Materials and Methods: A retrospective analysis was performed on patients with histopathologically confirmed FIGO stage II–III squamous cell carcinoma of the cervix treated with definitive IMRT or VMAT. Patient setup errors obtained from online kV-CBCT and MV-CBCT image guidance were retrieved from the treatment planning system and compared. Systematic and random errors were calculated, and age-based subgroup analysis (≤50 vs. >50 years) was performed.

Results

Eighteen patients were included in the analysis. No significant differences in setup errors were observed between kV-CBCT and MV-CBCT in the longitudinal, vertical, or lateral directions (all p>0.05). The two imaging modalities demonstrated moderate-to-good agreement intraclass correlation coefficient (ICC: 0.552–0.688), with minimal systematic bias on Bland–Altman analysis. Systematic and random setup errors were low, yielding van Herk planning target volume margins of approximately 3 mm in all translational directions. Although kV-CBCT detected a slightly higher frequency of setup deviations >3 mm, the difference was not significant. Patients aged ≥50 years exhibited significantly greater setup errors on MV-CBCT than those aged

Zitieren

APA

Mani, N., Singh, H. B., & Gupta, P. (2026). Individualized Nonadaptive Image Guided Radiotherapy (IGRT) In Stage II-III Carcinoma Cervix With Daily Cone Beam Computed Tomography (CBCT): Set Up Error Calculation. IOSR Journal of Dental and Medical Sciences, 25(7), 27-32. https://doi.org/10.9790/0853-2507012732

Vancouver

Mani N, Singh HB, Gupta P. Individualized Nonadaptive Image Guided Radiotherapy (IGRT) In Stage II-III Carcinoma Cervix With Daily Cone Beam Computed Tomography (CBCT): Set Up Error Calculation. IOSR Journal of Dental and Medical Sciences. 2026;25(7):27-32. doi:10.9790/0853-2507012732

BibTeX
@article{Mani2026,
  author = {Mani, Nilesh and Singh, Harikesh Bahadur and Gupta, Poonam},
  title = {Individualized Nonadaptive Image Guided Radiotherapy (IGRT) In Stage II-III Carcinoma Cervix With Daily Cone Beam Computed Tomography (CBCT): Set Up Error Calculation},
  journal = {IOSR Journal of Dental and Medical Sciences},
  year = {2026},
  volume = {25},
  number = {7},
  pages = {27--32},
  doi = {10.9790/0853-2507012732}
}
Kostenlos starten →