Normative Data and Repeatability of a 4-kHz Pure-Tone Psychoacoustic Tuning Curve Using a Single Clinical Audiometer
Journal of Audiology and Otology · 20.07.2026 · 30 (3) · 198-206
DOI: 10.7874/jao.2025.00675 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Audiology and Otology, erschienen am 20.07.2026. Verfasst von Byungwoo Yoon. Beteiligte Einrichtung: Department of Otorhinolaryngology, Busan St. Mary's Medical Center, Busan, Korea. wooya8207@gmail.com. Die Arbeit nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background and Objectives: This study aimed to establish normative data and evaluate the test-retest variability of 4 kHz pure-tone psychoacoustic tuning curves (PTCs) measured using a conventional two-channel clinical audiometer.Materials and Methods: Thirty-four adults with normal hearing (≤25 dB HL) underwent two PTC testing sessions separated by a 2-week interval. A 4,000-Hz signal tone presented at 10 dB above threshold and seven pure-tone maskers (3,150-4,896 Hz) were administered using an Interacoustics AC40 Audiometer through the 3-down/1-up approach. Participants were randomly assigned to either low-to-high or high-to-low masker-frequency-order groups. Test-retest variability was analyzed across masker frequencies, and correlations with participant demographics were determined. Q10 values were evaluated as indices of frequency selectivity.
Results
All participants completed the testing without auditory discomfort. Approximately 80% had a test-retest variability of ≤8 dB, indicating high reliability. The largest intersession differences were observed at 4,896 Hz (9.80±7.35 dB) and 3,150 Hz (5.77±6.95 dB). Test-retest variability was positively correlated with age (r= 0.50, p=0.006). The mean Q10 was 5.76±1.72, and decreased significantly with age (r=-0.49, p=0.014). The average tuning curve exhibited a V-shaped pattern with a tip at 4,000 Hz, consistent with classical findings.
Conclusions
The simplified pure-tone PTC protocol using a standard clinical audiometer provided reproducible and relevant estimates of auditory frequency selectivity. These findings support its clinical feasibility for assessing frequency selectivity and age-related changes in the auditory filter.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Otorhinolaryngology, Busan St. Mary's Medical Center, Busan, Korea. wooya8207@gmail.com.
Zitieren
APA
Yoon, B. (2026). Normative Data and Repeatability of a 4-kHz Pure-Tone Psychoacoustic Tuning Curve Using a Single Clinical Audiometer. Journal of Audiology and Otology, 30(3), 198-206. https://doi.org/10.7874/jao.2025.00675
Vancouver
Yoon B. Normative Data and Repeatability of a 4-kHz Pure-Tone Psychoacoustic Tuning Curve Using a Single Clinical Audiometer. Journal of Audiology and Otology. 2026;30(3):198-206. doi:10.7874/jao.2025.00675
BibTeX
@article{Yoon2026,
author = {Yoon, Byungwoo},
title = {Normative Data and Repeatability of a 4-kHz Pure-Tone Psychoacoustic Tuning Curve Using a Single Clinical Audiometer},
journal = {Journal of Audiology and Otology},
year = {2026},
volume = {30},
number = {3},
pages = {198--206},
doi = {10.7874/jao.2025.00675}
}
EUVIMED